
SIREMOBIL Compact
SPR2-130.061.02
Page 14 of 30
Siemens AG
System Manual
Rev. 11
09.04
CS PS 24
Medical Solutions
5 - 14
Adjustments / Programs
•
Close the status window by clicking on the OK button.
•
Respond to the query "Do you want to reset the unit" by clicking "No" .
•
Switch on fluoro.
•
Open the slot diaphragm to where the blades begin to disappear at the edge of the im-
age.
•
Switch off fluoro.
•
In the window "Available Functions:" select "FULL FORMAT POSITION" and click "Exe-
cute".
•
Close the status window by clicking on the OK button.
•
Respond to the query "Do you want to reset the unit" by clicking "No".
•
Select the zoom format.
•
Switch on fluoro.
•
Open the slot diaphragm to where the blades begin to disappear at the edge of the im-
age.
•
Switch off fluoro.
•
In the window "Available Functions:" select "ZOOM FORMAT POSITION" and click "Ex-
ecute".
•
Close the status window by clicking on the OK button.
•
Respond to the query "Do you want to reset the unit" by clicking "No" .
•
In the window "Available Functions:" select "Enable circle mask" and click "Execute"
•
Respond to the query "Do you want to reset the unit" by clicking "No" .
•
In the window "Available Functions:" select "INIT OFF & STORE POSITION" and click
"Execute".
•
Close the status window by clicking on the OK button.
•
Respond to the query "Do you want to reset the unit" by clicking "Yes". The adjustment
values are stored.
Concluding steps
5
•
Switch off the system and switch it back on again.
•
Return to the main menu by pressing ESC at the service PC.
•
Select the "Data, Backup..." menu.
•
Click "parameter" in the Backup window.
•
Then click "Backup". Enter "slot diaphragm adjusted" under "Remarks for Parameter".
•
Then click "OK" . The new position values are stored to disk.
•
You can now exit from the service program.