SIREMOBIL Compact
SPR2-130.061.02
Page 16 of 28
Siemens AG
System Manual
Rev. 11
09.04
CS PS 24
Medical Solutions
6 - 16
Replacing boards / replacing components
•
Configure the following parameters in User setup for MEMOSKOP:
•
Perform the IQ quick test.
Default Edge
Configure 1. Program a different Edge enhancement
value only upon customer request.
Monitor split
Configure Yes.
Monitor text
Configure Yes.
Disable dose rate selection in
user setup
Set to "No" if equipped with an area dose product mea-
surement device. Menu "Storage of Dose on Disk" and
Menu "Dose on Hardcopy output in User Setup are visi-
ble. Otherwise, set to "Yes".
Unit serial number
Program the serial number of the SIREMOBIL Compact.
Do not enter the first zero.
Product name
SIREMOBIL Compact (default). If wrong, configure
"SIREMOBIL Compact".
Default program changeable
Default: Yes.
Depending on country specific regulations, configure
Yes or No.
Software backward compatible Configure No.
Operating mode Parameters
Upon customer request, change the parameters pro-
grammed for the organ program.
VCR output
Configure to Monitor A.
Hardcopy Release
For SIREMOBIL Compact with 1 monitor:
Configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
Configure to monitor B.
Text monitor
For SIREMOBIL Compact with 1 monitor:
Configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
Configure to both monitors.
Split Screen
Configure horizontally. Configure vertically only upon
customer request.
Display of time
Configure Yes.
Set Date Time
Set the current date and time.
Hospital Name
Program the hospital name.
Storage Dose on Disk
If the area dose product meter is present, (Diamentor)
configure Yes. Otherwise, configure No.
Dose on hard copy output
If the area dose product meter is present, (Diamentor)
configure Yes. Otherwise configure No.