Installation and Start-Up
Checking the swivel-arm system
15
Rotary motion
15
1. System "OFF".
The rotary motion and the vertical adjustment of the swivel-arm system shall be
blocked.
2. System "ON", object table installed and projection angle set to minimum with
presetting knob (2/ Fig. 1).
3. Check the rotary motion of the swivel-arm system by pressing buttons (1/ Fig. 1)
one by one.
The rotation of the swivel-arm system shall stop when the button is released or
when end position is reached (+135 degrees clockwise and -180 degrees
counterclockwise). The projection angle is shown on the display at the lower part
of the stand.
Fig. 1 Swivel-arm system