M1000/3000 Nova
Register 3
SPB7-230.033.11
Page 2 of 6
Siemens AG
Installation and Start-Up
Rev. 06
05.06
CS PS SP
Medical Solutions
4 - 2
Installing the generator and the stand
Free-standing radiation shield (option)
4
For the installation of the free-standing radiation shield, please refer to separate instruc-
tions enclosed with the radiation shield.
Removing the generator cover
4
1. Pry loose the plastic strip (1) and remove the sixteen screws (2), eight on each
side.
2. Remove the front cover (3) from the generator.
Pay special attention to the contact washers (there are four con-
tact washers on either side).
They are needed again when reassembling the front cover to
establish protective ground connection and to fulfil EMC (Electro
Magnetic Compatibility) requirements.
Fig. 2 Removing the generator cover
NOTICE
3
1
2
M
A
M
0
0
0
9
1