M1000/3000 Nova
Register 3
SPB7-230.033.11
Page 18 of 22
Siemens AG
Installation and Start-Up
Rev. 06
05.06
CS PS SP
Medical Solutions
11 - 18
Calibrating and adjusting the AEC
16. When the exposure is done, the dialog
Calibration exposure
appears with values
for Dose rate and Diff. from nominal.
17. Save the values with <F2>.
18. Repeat steps 14 to 17 until all exposures are done.
19. When all exposures for one object table group are done, save the values with
<F2> in the dialog
Calibration object table group
.
20. The dialog
Calibration AEC correction table
appears. Select the next object table
group and repeat steps 5 to 19.
21. When all desired object table groups are calibrated, press <F2> to install the
calibrated correction table in the AEC.
If Er 013 occurs, acknowledge it and proceed.
If the Diff. from nominal exceeds
&
50% ensure that the exposure
really has taken place, the thickness of AEC calibration plexi are
correct and that the cassette (loaded with film) is inserted and
repeat the last exposure. If the difference still exceeds
+
50%
accept the value and proceed.
1 Help 2 Save
Exp support
3
4
5
6
7
8
9
10 Quit
M
A
M
0
0
7
6
6
<ESC> to exit, <F2> to save, <F3> to perform exposure
Mo/Mo H, Table ID: 1104, Object table group Grid
Exp.
5
Tension
25
kV
PMMA
4
cm
Dose
rate
267.23
Diff. from nominal
-21.47%
NOTE
NOTE
Installing correction table
ID: 1104 for Mo/Mo H
29% done
M
A
M
0
0
7
2
9
Calibration exposure