Appendix
22
-
7
Siemens AG
Register 3
SPB7-230.033.11
Page 7 of 12
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
AEC performance test
22
To be performed for the object tables, wings, and speeds in question.
D
OD
corr
= the difference between the two measured OD
corr
values in each case.
*Should error codes Er013 or Er450 appear; change to AEC calibration plexi which is
1 cm thinner.
If both speed H and speed D are to be tested, copy the test protocol.
Speed used for the tests below (H/D)
M1000 or Wing 1 M3000 Nova
22
Film
Screen
Manufacturer
Type
Manufacturer
Type
H
D
Grid Table 18x24 (Bucky)
mAs
OD
OD
below
OD
above
OD
corr
Mo/Mo 25 kV, 6 cm plexi*
Mo/Mo 32 kV, 2 cm plexi
Tolerances:
B
OD
corr
e
0.30 absolute
B
OD
corr
Grid Table 18x24 (Bucky)
mAs
OD
OD
below
OD
above
OD
corr
Mo/Rh 26 kV, 6 cm plexi*
Mo/Rh 32 kV, 2 cm plexi
Tolerances:
B
OD
corr
e
0.30 absolute
B
OD
corr
Grid Table 18x24 (Bucky)
mAs
OD
OD
below
OD
above
OD
corr
W/Rh
25 kV, 6 cm plexi*
W/Rh
32 kV, 2 cm plexi
Tolerances:
B
OD
corr
e
0.30 absolute
B
OD
corr