8 - 1
Siemens AG
Register 3
SPB7-230.033.11
Page 1 of 2
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
Checking the microprocessors
8
Microprocessors
8
The following LED displays on the processor P.C. boards indicate whether the relevant
microprocessors are operating correctly.
1. Mains and system "ON".
The following conditions must be indicated after not more than 15 s.
P.C. board D711:
• V9 "ON" (line voltage on).
• V13 "OFF" (system stand-by).
• V11 "ON" (system on).
• V16 "ON" (intermediate circuit - filament).
M
A
M
0
0
6
4
9
Microprocessors for:
LED displays
AEC: P.c. board D701
Master: P.c. board D702
Must be off.
Steadily on.
Must be off.
Flashing rapidly.
Must be steadily