M1000/3000 Nova
Register 3
SPB7-230.033.11
Page 6 of 10
Siemens AG
Installation and Start-Up
Rev. 06
05.06
CS PS SP
Medical Solutions
10
-
6
Checks
with
high
voltage
Oscilloscope diagrams
10
•
MAMMOMAT in mAs mode, see Fig. 3 - Fig. 10.
Checks indicated by
*
shall be disregarded in installations with
filter disc Mo 0.03 / Rh 0.025 (molybdenum rotating anode).
Fig.
3
Fig.
4
30 kV
1 V/T (1 T <=> 5 kV)
30 kV
1 V/T (1 T <=> 5 kV)
158 mA
1 V/T (1 T <=> 40 mA)
125 mA
1 V/T (1 T <=> 40 mA)
127
ms
20
ms/T
160
ms
20
ms/T
Record the measured kV and mAs value in the Test protocol, page
22 - 10.
NOTE
M
A
M
0
0
1
7
9
30 kV 20 mAs (P = 4.7 kW)
Tube P40 MoW,
MAMMOMAT 3000 Nova
* tungsten anode 0.3
Tube P40 MoW,
MAMMOMAT 1000/3000 Nova
molybdenum anode 0.3
M
A
M
0
0
1
1
2
30 kV 20 mAs (P = 3.75 kW)
NOTE