Calibrating and adjusting the AEC
11 - 17
Siemens AG
Register 3
SPB7-230.033.11
Page 17 of 22
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
13. Load the cassette with film. Since it will not be developed, the same film can be
used for all exposures during the following calibration.
14. To proceed with the calibration, place the cursor by one of the alternatives.
Press <F3>.
15. Follow the instructions in the message box shown.
1 Help 2 Save
Exp support
3
4
5
6
7
8
9
10
Quit
M
A
M
0
0
7
6
5
<ESC> to exit, <F2> to calibrate & save, <F3> to measure dose rate
Mo/Mo H, Table ID: 1104, Object table group Grid
Exp.
Plexi
kV
Dose rate
Diff. from nominal
1
3
25
---------
-------
2
3
28
---------
-------
3
4
25
---------
-------
4
4
32
---------
-------
5
4
35
---------
-------
6
6
25
---------
-------
7
7
28
---------
-------
8
7
32
---------
-------
9
7
35
---------
-------
Fig. 6 Dialog
Calibration object table group
shows the dose rates for a number of exposures. The dashed
lines below Dose rate indicate that no exposures have been performed yet.
Please put x cm of Plexi
on the object table.
Perform an exposure and
press <ENTER> when ready
or <ESC> to cancel.
<ENTER> - ok / <ESC> - cancel
M
A
M
0
0
7
2
8
Calibration object table group