M1000/3000 Nova
Register 3
SPB7-230.033.11
Page 12 of 22
Siemens AG
Installation and Start-Up
Rev. 06
05.06
CS PS SP
Medical Solutions
11 - 12
Calibrating and adjusting the AEC
Installation of AEC correction tables
11
1. In <Mainmenu> select <Configuration>, <AEC> and <AEC correction tables>.
The dashes in the ID column show that no correction table is installed (
Fig. 3
).
2. Make sure the backup floppy is inserted.
3. Decide for which combination you need to install correction table. (For available
identifications for correction tables, see Table 1 on Page 11-1 or the backup
floppy.)
4. Place the cursor by one of the desired Anode/Filter/Speed combinations.
Press <F2>.
If only one film/screen combination is to be used install
correction tables for H only.
1 Help 2 Inst
Cal
3
4
5
6
7
8
9
10
Quit
M
A
M
0
0
5
9
9
<ESC> to exit, <F2> to install new table, <F3> to calibrate existing
Anode/Filter
Speed
ID
Calibrated
Mag
Grid
No grid
Stereo
Mo/Mo
H
----
---
---
---
---
Mo/Mo
D
----
---
---
---
---
Mo/Rh
H
----
---
---
---
---
Mo/Rh
D
----
---
---
---
---
W/Rh
H
----
---
---
---
---
W/Rh
D
----
---
---
---
---
Existing tables in the AEC:
Fig. 3 AEC correction tables
NOTE
<F2> to continue, <ESC> to exit
New ID: 0000
M
A
M
0
0
6
0
0
AEC correction tables
Install new table