Cable
connections
5
-
11
Siemens AG
Register 3
SPB7-230.033.11
Page 11 of 14
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
Connecting the incoming mains to the stand (400 V, 2-phase)
5
400 V 2-phase connection
5
1. Connect the incoming mains cable (1) to X899 (2) in the stand as shown in Fig. 13
and Fig. 14.
2. Fit plastic tubes (4) on the plates.
3. Strain-relieve (3) the incoming mains cable.
Fig. 13 Incoming mains cable
Fig. 14 Incoming mains, 2-phase
1
M
A
M
0
0
0
9
7
L
1
L
2
L
3
N
1
2
3
1
4
1
TWO-PHASE
GENERATOR
STAND
F10
F20
X881
Incoming mains
208, 230, 240, 277 or 400 VAC
P.E N
L1 L2 L3
X899
M
A
M
0
0
0
9
5