Appendix
22
-
5
Siemens AG
Register 3
SPB7-230.033.11
Page 5 of 12
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
Starting up and using the service PC - Dose Calculation program
22
1. Switch on generator and service PC.
After initialization, the service PC shows:
C:\>
2. Now insert the diskette with the dose calculating program.
3. Select the appropriate drive
and then press <ENTER>. The screen shows:
A:\>
or
B:\>
4. Start the service program by typing "DOSE_CFG", then press <ENTER>.
The program asks for the user’s name:
Your name, please
Type the name of the technician, for example NN, and then press <ENTER>.
5. The program asks for the password:
Password, please
Type the password (********) and then press <ENTER>. This is the same
password that is used for the regular service program.
6. The display window shows:
PROGRAM-MODE; Mode: Normal
If
Mode: Normal
is the correct choice then press <ENTER> or
if
Mode: Stand-alone
is wanted press the <SPACE> key once and then
press <ENTER>.
The display window shows:
Main menu
7. Select the program part to be used with keys <
↑>
and/or <
↓>
, then press
<ENTER>.
8. Make the necessary entries in the appropriate part of the program. Save the
entered data with <F2>. Page back in the program with <ESC>. The appropriate
instructions are shown on the monitor.
End the procedure with the service PC with <F10>.
Troubleshooting
22
If communication with the MAMMOMAT still cannot be established, follow the instructions
in this section. The following procedure will disable the buffering of the RS232 port. (This
example is for Windows
®
2000 Pro. Windows
®
95 and Windows
®
98 has a similar
approach, consult the OS documentation for details.)
1. Choose Settings from the Start Menu.
2. Choose Control Panel.
3. Choose System.
4. Choose the tab Hardware.
5. Press the button Device Manager.
6. Expand the row Ports.
7. Double-click on the COM1 row.
8. Choose the tab Port Settings.
9. Press the button Advanced.
10. Uncheck the check box Use FIFO Buffers.
11. Exit the dialog by pressing the OK-button.
{A:}
{B:}