Checks
with
high
voltage
10
-
3
Siemens AG
Register 3
SPB7-230.033.11
Page 3 of 10
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
6. If necessary, adjust R43 on D702 (Fig. 2) and make another exposure. Note the
new values in the kV-adjustment protocol.
7. Mains and system OFF.
8. Disconnect DVM from R72 and R73.
Do not overload the X-ray tube by making too many exposures in a
short time.
R43 does not need to be relocked with locking paint.
Decrease
kV
Increase
kV
Potentiometer
R43
M
A
M
0
0
5
2
8
Fig. 2
NOTE