STATUS MESSAGES
9.2
Alarm messages
Alarm messages not listed: Contact one of the SICK service points.
9.3
Maintenance messages
Error message
Relay contact
Description
Reference air flow too low
Possible causes:
Reference air flow rate incorrectly set, instrument air supply
to the system insufficient, reference air pump defective.
Reference air flow too
high
Possible cause:
Reference air flow rate incorrectly set.
Limit alarm 1
O
2
limit value alarm 1,
open
Possible cause:
The O
2
measured value underflows/overflows the specified O
2
alarm limit.
Limit alarm 2
O
2
limit value alarm 2,
open
Possible cause:
The O
2
measured value underflows/overflows the specified O
2
alarm limit.
Electronic temp. too low
Possible cause:
The ambient temperature of the control unit falls below the
specified lower limit. The measured value tolerances specified for the system
are no longer guaranteed.
Electronic temp. too high
Possible cause:
The ambient temperature of the control unit falls below the
specified lower limit. The measured value tolerances specified for the system
are no longer guaranteed.
Clock battery low
This cannot can be reset by the user; it is only reset automatically after the
watch battery (lithium cell 2032) has been replaced. The alarm has no effect
as long as the system is supplied with power voltage. The set time/date can be
incorrect after switching the system off and on again. Any timer-controlled
automatic adjustment that may have been set can then no longer be carried
out correctly.
The battery may only be replaced by SICK.
Maintenance message:
Relay contact
Description
Measured value(s) held
Maintenance,
closed
When the measured value memory is switched on, the O
2
measured value
determined before an adjustment is held at the mA output for the duration of
the status message.