O P E R A T I N G I N S T R U C T I O N S | ZIRKOR200 Ex-G
Subject to change without notice
7
MENU OVERVIEW AND EXPLANATIONS
7.2.10
REMOTE
The password is used:
●
For authentication and pairing with a smartphone/tablet/notebook/PC.
●
For authentication/login after each connection establishment. Without authentication/
login, neither device data can be read out nor the device configuration changed.
The
range
limits the transmission power of the REMOTE module.
Maximum
= approx.
100 m,
medium
= approx. 10 m,
short
= approx. 1 m. The actual possible range can vary
greatly due to local conditions and the reception performance of the smartphone used.
Fig. 29: REMOTE connection active
If a SICK-REMOTE connection to the analyzer is active, the connection is shown in the
upper right corner of the display (1).
7.2.11
Units of measure
Sets the units of measurement for temperature (°C / °F) and pressure (psi / mbar).
7.2.12
Language
Sets the language for all display texts. The languages German, English, Spanish, Polish and
French are available.
Note:
REMOTE is disabled by default, so the password and range are not shown. After
activation (only possible from system level), the 8-digit password which is prompted
during the remote connection setup must be assigned.
HINTS:
A maximum of 16 users (smartphones / tablets) can be connected to the REMOTE
module of a SICK analyzer.
The connection fails when additional users try to connect. In this case, REMOTE must
be switched off and on again via the front panel, all mobile devices that have already
been connected to the device once must be registered to it again.