COMMISSIONING
6.4
Initial commissioning
Before operating the unit, make sure that:
1 The enclosure is tightly closed.
2 The device power supply is activated at an external location (e.g. main switch).
After switching the device on, the startup screen which also contains the software version
is displayed briefly. Afterwards, you will be prompted to select the language, set the system
date and time, assign a TAG number and select a SICK REMOTE code (only if SICK REMOTE
is activated at the factory).
The measuring probe heating phase begins and then measuring mode starts.
Fig. 23: Initial commissioning. The software version is displayed in the lower right corner.
6.5
Display - heating process
Fig. 24: Display - heating process
6.6
Display - Measuring mode
Fig. 25: Display - Measuring mode
* Only when O
2
limit value alarms are switched on and the defined limit values are within the
measuring range.
The heating process starts with heating up the
O
2
sensor.
1
TAG No.
2
Temperature, analog display
3
Last access, with date
Rising measuring probe
temperature
4
(or) shows waiting time
(or) heating fault
5
Current temperature
6
Softkey: System menu
.
1
TAG No.
2
Measured value
3
Measuring range specification
4
Flashing arrow - indicators show measuring
range underflow or overflow
5
Last access, with date
6
Measuring component
7
Analog display shows measured value and
range indicator - min-alarm / max-alarm *
8
Softkey name