O P E R A T I N G I N S T R U C T I O N S | ZIRKOR200 Ex-G
Subject to change without notice
5
INSTALLATION
Fig. 13: Connection cable
5.9.9
Fitting the pneumatic unit
▸
Mount on the wall.
▸
Make the pneumatic connections.
▸
Connect connection cable for solenoid valve (observe the manufacturer's Instructions).
▸
Connect grounding from the outside via a PE connection (M8x15).
▸
Firmly install the cable between pneumatic unit and control unit.
Probe cable
No. Function
Diameter
Colors
Additional
information
1
Measuring cell
2 x 0.75 mm² White-brown / brown
With shielding
2
Thermoelement
2 x 0.75 mm² Green / white
With shielding
3
Thermoelement
2 x 0.75 mm² Green / white
With shielding
4
Measuring probe
heater
3 x 1.5 mm²
Black / blue / yellow-green
5
Solenoid valve of the
analyzer unit
2 x 0.75 mm² Grey / grey-blue
Without function for
ZIRKOR200 Ex-G
6
Shielding
7
Not used
1 1 2 2
1 2 3 4
1
2
3
4
5 6
7