INSTALLATION
5
5.10.7
Connecting the connection cable to the control unit
5.10.8
Connecting the valve cable of the analyzer unit to the control unit
5.10.9
Connecting the pneumatic unit line to the control unit
5.10.10
Outputs/functions and relay assignment
All relay contacts are designed for 24 V and 1 A ~, 1 A = (exception: Probe valve).
* The system error relay is also active during the heating phase.
** The relay contact of the analyzer unit valve is designed for 230 V and 1 A
.
Connection
terminal
Function
Wire bundle
No.
Cross-section
(mm²)
Colors
Additional
information
TERM.X3:6
Measuring
probe heater
3
1.5
Black
TERM.X3:8
Measuring
probe heater
3
1.5
Blue
TERM.X3:10
Measuring
probe heater
3
1.5
Green-yel-
low
TERM.X3:11
Measuring
chamber
1
0.75
Brown
With shielding
TERM.X3:12
Measuring
chamber
1
0.75
White
With shielding
X2.3
Thermoelement 2a
0.75
Green
With shielding
X2.4
Thermoelement 2a
0.75
White
With shielding
X2.5
Thermoelement 2b
0.75
Green
With shielding
X2.6
Thermoelement 2b
0.75
White
With shielding
Connection
terminal
Function
Wire bundle
No.
Cross-section
(mm²)
Colors
Additional
information
PWRB.X5:24A Solenoid valve
1
Black
PWRB.X5:24B Solenoid valve
1
Blue
X4.5
Solenoid valve
1
Green-yellow
Connection
terminal
Function
Wire bundle
No.
Cross-section
(mm²)
Colors
Additional
information
X2.3
Solenoid valve
1
Black
X2.4
Solenoid valve
1
Blue
X4.4
Solenoid valve
1
Green-yellow
Connection
terminal
Relay output
Contact
Function
PWRB.X5:19A/B System error*
NC contact
Signals operation-critical errors
PWRB.X5:18A/B Maintenance NO
contact
System code was entered, system in
Maintenance mode.
PWRB.X5:20A/B Measuring range
NO contact
Closed: Measuring range 1 active
Open: Measuring range 2 active
PWRB.X5:24A/B Analyzer unit
valve**
NO contact
Control of the analyzer unit valve
PWRB.X5:21A/B Limit value 1
NC contact
Signal an O
2
limit value exceeds 1
st
limit value
PWRB.X5:22A/B Limit value 2
NC contact
Signal an O
2
limit value exceeds 2
nd
limit value