O P E R A T I N G I N S T R U C T I O N S | ZIRKOR200 Ex-G
Subject to change without notice
INSTALLATION
5
SICK recommends fitting the analyzer unit horizontally (-1 ° bis - 3 °) to ensure the fastest
possible response time.
5.9.6
Aligning the V-shield
Before fitting the analyzer unit, determine flow direction (1) of the flue gas and turn V-shield
(2) of the filter head against the exhaust gas flow. Filter head (3) is freely rotatable by 360°
on a thread. To do this, completely loosen hexagon socket screws (4) on the filter head,
align the filter head with V-shield with the help of jaw wrench (5) and tighten the hexagon
socket screws again.
Observe ambient temperature on the connection
unit and process temperature.
Avoid vibrations greater than 2 g.
Observe protection class.
Heavy device, ensure correct lifting and carrying.
WARNING: Danger of explosion through breaching the flameproof enclosure
An Allen screw, damaged due to incomplete loosening or during filter replacement or a
loose Allen screw can damage the flameproof enclosure by damaging the flameproof
joint and thus lead to an explosion.
▸
Loosen and tighten the hexagon socket screw completely.
▸
Replace damaged hexagon socket screws.
WARNING: Danger of explosion due to loose filter head
Ensure that the filter head is completely screwed on to the thread (to the stop). From
this end position, the filter head may be turned back once by max. 360° to align the
filter head.
Explosion protection can no longer be guaranteed when the filter head is unscrewed
further than 360°.