O P E R A T I N G I N S T R U C T I O N S | ZIRKOR200 Ex-G
Subject to change without notice
INSTALLATION
5
▸
For version with additional signal connections (option): Observe the individual
information provided.
▸
Connect the signal cables to the desired connection terminals in the lower enclosure
section.
▸
For version with additional signal connections (option): Observe the individual
information provided.
DANGER:
Electrostatic discharges can destroy electronic components and there is a danger of fire
and explosion.
▸
Before contact with electrical connections and internal components: Ground the
human body and the tool used to discharge electrostatic charges.
Recommended method:
▸
When the protective conductor is connected: Touch a blank metal part of the
enclosure.
▸
Otherwise: Touch a different blank metal surface that is connected to the protective
conductor or has safe contact to the grounding.
▸
Pay primary attention to any individual information provided.
No.
Connection
terminals
Polarity
Function
17
PWRB.X5:17A +
Current output – 0
2
0/4 ... 20 mA
PWRB.X5:17B -
18
PWRB.X5:18A +
Maintenance - Active = closed
PWRB.X5:18B -
19
PWRB.X5:19A +
System error - Active = open
PWRB.X5:19B -
20
PWRB.X5:20A +
Measuring range - MBI = closed
PWRB.X5:20B -
21
PWRB.X5:21A +
Limit value 1 - Active = open
PWRB.X5:21B -
22
PWRB.X5:22A +
Limit value 2 - Active = open
PWRB.X5:22B -
23
PWRB.X5:23A +
Solenoid valve of the analyzer unit – power supply
PWRB.X5:23B -
24
PWRB.X5:24A +
Solenoid valve of the analyzer unit – outlet
PWRB.X5:24B -
25
PWRB.X5:25A +
Measuring range – 12 ... 24 V DC
PWRB.X5:25B -
26
PWRB.X5:26A +
NC
PWRB.X5:26B -
27
PWRB.X5:27A +
Calibration release – 12 ... 24 V DC
PWRB.X5:27B -
28
PWRB.X5:28A +
NC
PWRB.X5:28B -
29
PWRB.X5:29A +
NC
PWRB.X5:29B -
30
PWRB.X5:30A +
NC
PWRB.X5:30B -