O P E R A T I N G I N S T R U C T I O N S | ZIRKOR200 Ex-G
Subject to change without notice
5
INSTALLATION
5.10.14
Connecting the analyzer electronics to the connection cable
Fig. 18: Electrical connections on the analyzer unit
Thread adapter 1“NPT to ¾“ NPT for ATEX approved cable gland 3/4“ NPT
Connection
terminal
Function
Wire bundle
No.
Cross-section
(mm²)
Colors
Additional
information
1
Measuring cell
1
0.75
White (br)
Do not connect the
shielding
2
Measuring cell
1
0.75
Brown
Do not connect the
shielding
3
Thermocouple 1 2a
0.75
Green
Do not connect the
shielding
4
Thermocouple 1 2a
0.75
White
Do not connect the
shielding
31
Thermocouple 2 2b
0.75
Green
Do not connect the
shielding
41
Thermocouple 2 2b
0.75
White
Do not connect the
shielding
6
Measuring
probe heater
3
1.5
Black
7
Measuring
probe heater
3
1.5
Blue
8
Measuring
probe heater
3
1.5
Green-yellow
Note:
Under no circumstances should the shielding of the connection cable be laid to the
probe side
, as this will lead to measurement inaccuracies.