LCR-Reader-MP Manual
Siborg Systems Inc.
22
0.65 Vrms
0.5%+2
0.2%+2
0.2%+2
0.5%+2
1MΩ
0.1кΩ
0.1Vrms
2%+5
2%+5
2%+5
3%+5
Parallel
0.65 Vrms
1.0%+3
0.5%+3
0.5%+3
5%+3
10MΩ
0.001MΩ
0.1Vrms
3%+5
3%+5
5%+5
-
Parallel
0.65 Vrms
2%+5
2%+5
3%+5
-
20MΩ
0.01MΩ
0.1Vrms
-
-
-
-
Parallel
0.65 Vrms
3%+5
3%+5
-
-
•
Calibrate with short and open test leads before measurement
•
If D exceeds 0.1 then multiply the result by
2
1
D
•
Measurement time 1 sec
14.4.3
DC Resistance
Range
Resolution
Accuracy
10Ω
0.001Ω
1%+20
100Ω
0.01Ω
0.5%+3
1кΩ
0.1Ω
0.3%+2
10кΩ
0.001кΩ
0.3%+2
100кΩ
0.01кΩ
0.3%+2
1MΩ
0.1кΩ
0.5%+3
10MΩ
0.001MΩ
3%+5
20MΩ
0.01MΩ
5%+5
•
Calibrate with short and open test leads before measurement
•
Test signal 1.2 V
14.4.4
Capacitance
Range
Resolution
Level
Test Frequency
50Hz
100Hz, 120Hz
1kHz
10kHz
20kHz-75kHz
100kHz
10 pF
0.00 1pF
0.1Vrms
-
-
-
-
-
-
0.65 Vrms
-
-
-
-
-
5%+50*
100 pF
0.01 pF
0.1Vrms
-
-
-
-
-
-
0.65 Vrms
-
-
-
3%+5*
3%+5*
2%+5*
1000 pF
0.1 pF
0.1Vrms
-
-
-
5%+5*
5%+5*
5%+5*
0.65 Vrms
-
-
5%+10*
0.3%+2*
0.5%+2*
0.5%+2*
10 nF
0.001 nF
0.1Vrms
-
-
5%+5
2%+5
2%+5
2%+5
0.65 Vrms
-
5%+20
0.3%+2
0.3%+2
0.5%+2
0.5%+2
100 nF
0.01 nF
0.1Vrms
-
5%+5
2%+5
2%+5
2%+5
2%+5
0.65 Vrms
-
0.5%+2
0.3%+2
0.3%+2
0.5%+2
0.5%+2
1000 nF
0.1 nF
0.1Vrms
-
2%+5
2%+5
2%+5
3%+5
3%+5
0.65 Vrms
-
0.5%+2
0.3%+2
0.3%+2
0.5%+3
0.5%+3
10
μF
0.001
μ
F
0.1Vrms
-
2%+5
2%+5
3%+5
-
-
0.65 Vrms
3%+3
0.5%+2
0.3%+2
0.3%+2
3 %+5
3 %+5
100
μ
F
0.01
μ
F
0.1Vrms
5%+5
2%+5
3%+5
-
-
-
0.65 Vrms
2%+2
0.5%+2
0.3%+2
3 %+5
-
-