Insert the test leads at the distance of the size of the component and calibrate with open probes. Only
then measure the capacitor.
Before measuring small inductances (less than 1
μ
H) and resistances (less than 10 Ohm) wipe off the tips
of any dirt and oxides and calibrate with short probes.
7.1.4
Offset Calibration Board
The Offset Calibration Board provides a reliable method of determining the parasitic offset
between the tips. The dummy PCB uses holes to represent various sizes of components.
To use the board, place the tips into the corresponding holes to the size of component under
test; set open/short calibration by turning the jog-wheel clockwise and hold for 2 beeps and
release. The result is the actual value of the component.
7.2
R-L-C-D mode menu
7.2.1
Primary Parameter
In Auto mode the type of measured component is determined automatically: R: Resistor, L: Inductance,
C: Capacitor, D: Diode. Measurement mode can also be set manually to R, L, C, Z as the impedance.