LCR-Reader-MP Manual

 

Siborg Systems Inc. 

13 

 

In order to set the default measurement mode when the device is turned on, select the mode, press the 
jog-wheel and hold for 2 beeps. This measurement setting will be 

stored in the device’s non

-volatile 

memory for all other measurements until selecting other parameters or Default in the menu. 

7.2.10 

Large Cap 

In this mode, capacitance and ESR of capacitors from 1 

μF

 to 40 mF are measured. The capacitance is 

measured at a frequency of 50Hz, and the capacitor's ESR is measured at the frequency of 100 kHz. 

Discharge Capacitors before testing! 

If open p-n junctions of semiconductor components cause an error in the measurements, then the 
screen shows the diode symbol and the direction of the p-n junction is shown instead of the primary 
parameter type. 

If the resistance of the component is less than 10 ohms, there will be short beeps. Before the first 
application in this sub-mode it is necessary to calibrate the device with both open and shorted probes. 

For the rejection of electrolytes based on ESR, two tables are stored in the memory of the device, one 
for standard aluminum capacitors and the other for 

Low ESR

 capacitors. Two additional numbers are 

displayed on the screen in this case: the first number is the quality of the capacitor. For standard 
aluminum capacitors, positive values indicate rejection, and for 

Low ESR

 rejected values are above -10. 

The second number is selected operating voltage of the capacitor in menu under C voltage: 6.3, 10, 16, 
2, 3, 50-63, 100, and 160+ Volts. This menu opens by clockwise turn of the jog-wheel. 

7.2.11 

Large Cap (1 uF to 40 mF) Settings 

 

Clockwise for 1 beep to open the mode menu. 

 

Clockwise for 2 beeps to make calibration eliminating the offset. Only short calibration is 
required. 

 

Counter clockwise for 2 beeps to toggle the test signal level between 0.65 Vrms and 0.1 
V peak-to-peak. 

7.2.12 

Super Large Cap > 40 mF Measurement 

• 

When in the Large Cap mode, counter clockwise for 1 beep to switch to DC capacitance 

measurement mode for Capacitors larger than 40 mF up to 680 mF. To speed up the settling 
of the readings, short the capacitor leads for 2-3 seconds before the measurements. Counter 
clockwise for 1 beep again to return to the normal Large Cap mode.  

 

Содержание LCR-Reader-MP

Страница 1: ......

Страница 2: ...ing off the device 7 6 3 Extended Voltage Range Switch 7 6 4 Controlling device using its orientation 7 6 5 Charging the Battery 7 6 6 Main and System Menu 8 7 Modes of Operation 9 7 1 R L C D Mode 9...

Страница 3: ...mode settings 16 7 6 1 Controls 16 7 7 Frequency Meter mode 16 7 7 1 Changing the mode 16 7 7 2 Frequency Period control 16 7 8 Signal Generator Mode 17 7 8 1 Controls 17 8 Default Settings 18 8 1 De...

Страница 4: ...tion 21 14 4 Details of the Measurement Modes 21 14 4 1 Test Signal Generator 21 14 4 2 Resistance 21 14 4 3 DC Resistance 22 14 4 4 Capacitance 22 14 4 5 DC Capacitance Measurement 23 14 4 6 Inductan...

Страница 5: ...aphics for alternating voltage measurements Jog dial control wheel Multiple Parameter Display primary secondary etc Screen orientation Automatic Manually Sound indication Battery discharge warning ful...

Страница 6: ...onal to the current flowing through the measured component After digitizing the ADC signals the impedance is calculated according to the formula DUT impedance Z Rj Vau Vaj Values of Impedance obtained...

Страница 7: ...the number of pulses from the reference generator and fr is the frequency of the reference generator The principle of measuring the voltage is based on comparing the input signal with the reference vo...

Страница 8: ...the device perpendicular to the floor the screen will flip orientation to the hand being used 2 Turn the probes up to turn on off the indicator light 3 If you place the device on a flat surface with...

Страница 9: ...item by moving the jog wheel in a clockwise or counter clockwise direction and pressing down to select To select the menu item highlight the corresponding menu item and press the jog wheel To set the...

Страница 10: ...the test leads the screen displays information depending on the component and the selected settings in the R L C D mode menu Capacitance Inductance Resistance Diode In automatic mode measurement of c...

Страница 11: ...parasitic capacitance is obtained Please use LCR Reader Calibration Board to set a proper distance between the test leads according to the component size because the parasitic capacitance depends on t...

Страница 12: ...able method of determining the parasitic offset between the tips The dummy PCB uses holes to represent various sizes of components To use the board place the tips into the corresponding holes to the s...

Страница 13: ...components and in circuit measurements of R C L When measuring on boards if at high test signal level opens p n junctions of semiconductors components causing an error in the measurements then the si...

Страница 14: ...electrolytes based on ESR two tables are stored in the memory of the device one for standard aluminum capacitors and the other for Low ESR capacitors Two additional numbers are displayed on the scree...

Страница 15: ...diodes the polarity of the diode the voltage drop in the open state and the reverse current Ir are displayed Diode mode is used for diode parameter extraction using DC bias It is recommended for in ci...

Страница 16: ...to a current and displayed on the screen In the upper line of the screen the resistance is displayed taking into account the input resistance of the device and the lower value of the current To enter...

Страница 17: ...for 2 beeps to automatic selection of T pix V div Press and hold up to 2 beeps to enter the waiting oscilloscope mode In the window set the parameters T pix and V div as indicated above Press and hold...

Страница 18: ...ty cycle and the waveform icon The frequency in Hz is displayed in the center of the screen The self shutdown time is 4 times longer than the default in other modes An external decoupling capacitor is...

Страница 19: ...rms Measurement Period 0 5 s Equivalent Circuit Auto Sound On Data Hold Off 8 2 Default Settings in System Menu The following settings are stored in the device EEPROM when Default in the System Menu i...

Страница 20: ...ht c Use a soft cloth dampened with water to clean the outside surface and cleaning the LCD display device d Do not use liquid solvents or detergents 10 2 Repairs If there is an unexpected measurement...

Страница 21: ...of operation 14 1 Overview R L C Measurement Parameters C R L R R C D C ESR L Q Z Measuring Time 0 25 s 0 5 s 1 0 s 2 0 s Generator Test Signal Frequency 50 Hz 100 Hz 120 Hz 1 kHz 10 kHz 20 kHz 30 kHz...

Страница 22: ...x80x25mm Unit Weight 45 g Weight of the Device and Case 103 g 14 4 Details of the Measurement Modes 14 4 1 Test Signal Generator Test Frequency 50Hz 100Hz 120Hz 1kHz 10kHz 20kHz 30kHz 40kHz 50kHz 60kH...

Страница 23: ...3 5 20M 0 01M 5 5 Calibrate with short and open test leads before measurement Test signal 1 2 V 14 4 4 Capacitance Range Resolution Level Test Frequency 50Hz 100Hz 120Hz 1kHz 10kHz 20kHz 75kHz 100kHz...

Страница 24: ...rement Measurement time 1 sec 14 4 5 DC Capacitance Measurement Range Resolution Error 40 100mF 0 01mF 5 5 680mF 0 1mF 5 5 Discharge capacitors before measurement 14 4 6 Inductance Range Resolution Le...

Страница 25: ...ut Impedance of the Instrument Frequency 75 1000 mV 0 1 mV 5 3 1 kOhm 30 50000 Hz 1 0 1 2 V 0 001 V 5 3 1 kOhm 30 50000 Hz 0 1 15 V 0 01 V 5 3 10 mOhm 30 3000 Hz Presence of the DC offset 5mV will res...

Страница 26: ...0 0 86 0 72 0 57 0 50 0 43 0 36 0 22 470 0 71 0 61 0 50 0 40 0 35 0 30 0 25 0 16 560 0 59 0 51 0 42 0 34 0 30 0 25 0 21 0 13 680 0 49 0 42 0 35 0 28 0 24 0 21 0 17 1000 0 33 0 29 0 24 0 19 0 17 0 14...

Страница 27: ...85 330 0 25 0 17 0 15 0 11 0 10 0 091 0 068 470 0 18 0 14 0 093 0 088 0 084 0 074 1000 0 066 0 063 0 060 0 057 0 054 0 048 2200 0 038 0 036 0 034 0 032 0 031 0 027 3300 0 032 0 030 0 029 0 027 0 026 4...

Отзывы: