background image

LCR-Reader-MP Manual

 

Siborg Systems Inc. 

10 

 

In this case two additional figures are displayed. The first number is the quality of the capacitor. For 
standard aluminum capacitors a positive number indicate rejection, and for 

Low ESR

 values above -10. 

The 2nd number is the selected operating voltage of the capacitor. It can be specified in the menu item 

Large Cap/C-voltage

 

 

 

Discharge Capacitors before testing! 

 

7.1.2 

Quick Controls 

 

Turn the jog-wheel clockwise* and hold for 1 beep to change the test parameters (more info 

below) 

 

Turn the jog-wheel counter clockwise and hold for 1 beep to change the test frequency. The 

frequency changes cyclically Auto-100Hz-120Hz-1kHz-10kHz-100kHz. If the frequency is selected 
from the menu in the range 20-75kHz, the frequency changes cyclically 20kHz-30kHz-40kHz-
50kHz-60kHz-75kHz-100kHz  

 

Turn the jog-wheel clockwise and hold for 2 beeps to eliminate the offset. If the probes are 
shorted, parasitic resistance is obtained. If the probes are open, parasitic capacitance is 
obtained. Please use LCR-Reader Calibration Board to set a proper distance between the test 
leads (according to the component size) because the parasitic capacitance depends on the 
distance. 

 

Turn the jog-wheel counter-clockwise and hold for 1 beep to toggle the signal level between 
Auto and fixed 0.65Vrms 

 

If the probes are not connected to anything, press and hold up to 2 beeps to toggle the test 
frequency between Auto and 100 kHz. If a part is connected to the probes, then relative 
measurements are made showing the difference from the first measured component. Additional 
symbol in the top-middle of the screen is displayed. 

 
 

*When connected to the Capacitance is greater 
than 1 

μ

F range for the operating voltage 

 

 

7.1.3 

Open/Short Calibration 

Before the first use, the device must be calibrated with Open and Short probes for each test frequency. 
When the device is set to Auto-frequency the device automatically calibrates at 100Hz, 1 kHz, 10kHz, 
and 100kHz. 

 To eliminate the offset, turn the jog-wheel clockwise, hold for 2 beeps and release. When measuring 
small capacitances less than 100 pF, use LCR-Reader Offset Calibration Board included in the package. 

Содержание LCR-Reader-MP

Страница 1: ......

Страница 2: ...ing off the device 7 6 3 Extended Voltage Range Switch 7 6 4 Controlling device using its orientation 7 6 5 Charging the Battery 7 6 6 Main and System Menu 8 7 Modes of Operation 9 7 1 R L C D Mode 9...

Страница 3: ...mode settings 16 7 6 1 Controls 16 7 7 Frequency Meter mode 16 7 7 1 Changing the mode 16 7 7 2 Frequency Period control 16 7 8 Signal Generator Mode 17 7 8 1 Controls 17 8 Default Settings 18 8 1 De...

Страница 4: ...tion 21 14 4 Details of the Measurement Modes 21 14 4 1 Test Signal Generator 21 14 4 2 Resistance 21 14 4 3 DC Resistance 22 14 4 4 Capacitance 22 14 4 5 DC Capacitance Measurement 23 14 4 6 Inductan...

Страница 5: ...aphics for alternating voltage measurements Jog dial control wheel Multiple Parameter Display primary secondary etc Screen orientation Automatic Manually Sound indication Battery discharge warning ful...

Страница 6: ...onal to the current flowing through the measured component After digitizing the ADC signals the impedance is calculated according to the formula DUT impedance Z Rj Vau Vaj Values of Impedance obtained...

Страница 7: ...the number of pulses from the reference generator and fr is the frequency of the reference generator The principle of measuring the voltage is based on comparing the input signal with the reference vo...

Страница 8: ...the device perpendicular to the floor the screen will flip orientation to the hand being used 2 Turn the probes up to turn on off the indicator light 3 If you place the device on a flat surface with...

Страница 9: ...item by moving the jog wheel in a clockwise or counter clockwise direction and pressing down to select To select the menu item highlight the corresponding menu item and press the jog wheel To set the...

Страница 10: ...the test leads the screen displays information depending on the component and the selected settings in the R L C D mode menu Capacitance Inductance Resistance Diode In automatic mode measurement of c...

Страница 11: ...parasitic capacitance is obtained Please use LCR Reader Calibration Board to set a proper distance between the test leads according to the component size because the parasitic capacitance depends on t...

Страница 12: ...able method of determining the parasitic offset between the tips The dummy PCB uses holes to represent various sizes of components To use the board place the tips into the corresponding holes to the s...

Страница 13: ...components and in circuit measurements of R C L When measuring on boards if at high test signal level opens p n junctions of semiconductors components causing an error in the measurements then the si...

Страница 14: ...electrolytes based on ESR two tables are stored in the memory of the device one for standard aluminum capacitors and the other for Low ESR capacitors Two additional numbers are displayed on the scree...

Страница 15: ...diodes the polarity of the diode the voltage drop in the open state and the reverse current Ir are displayed Diode mode is used for diode parameter extraction using DC bias It is recommended for in ci...

Страница 16: ...to a current and displayed on the screen In the upper line of the screen the resistance is displayed taking into account the input resistance of the device and the lower value of the current To enter...

Страница 17: ...for 2 beeps to automatic selection of T pix V div Press and hold up to 2 beeps to enter the waiting oscilloscope mode In the window set the parameters T pix and V div as indicated above Press and hold...

Страница 18: ...ty cycle and the waveform icon The frequency in Hz is displayed in the center of the screen The self shutdown time is 4 times longer than the default in other modes An external decoupling capacitor is...

Страница 19: ...rms Measurement Period 0 5 s Equivalent Circuit Auto Sound On Data Hold Off 8 2 Default Settings in System Menu The following settings are stored in the device EEPROM when Default in the System Menu i...

Страница 20: ...ht c Use a soft cloth dampened with water to clean the outside surface and cleaning the LCD display device d Do not use liquid solvents or detergents 10 2 Repairs If there is an unexpected measurement...

Страница 21: ...of operation 14 1 Overview R L C Measurement Parameters C R L R R C D C ESR L Q Z Measuring Time 0 25 s 0 5 s 1 0 s 2 0 s Generator Test Signal Frequency 50 Hz 100 Hz 120 Hz 1 kHz 10 kHz 20 kHz 30 kHz...

Страница 22: ...x80x25mm Unit Weight 45 g Weight of the Device and Case 103 g 14 4 Details of the Measurement Modes 14 4 1 Test Signal Generator Test Frequency 50Hz 100Hz 120Hz 1kHz 10kHz 20kHz 30kHz 40kHz 50kHz 60kH...

Страница 23: ...3 5 20M 0 01M 5 5 Calibrate with short and open test leads before measurement Test signal 1 2 V 14 4 4 Capacitance Range Resolution Level Test Frequency 50Hz 100Hz 120Hz 1kHz 10kHz 20kHz 75kHz 100kHz...

Страница 24: ...rement Measurement time 1 sec 14 4 5 DC Capacitance Measurement Range Resolution Error 40 100mF 0 01mF 5 5 680mF 0 1mF 5 5 Discharge capacitors before measurement 14 4 6 Inductance Range Resolution Le...

Страница 25: ...ut Impedance of the Instrument Frequency 75 1000 mV 0 1 mV 5 3 1 kOhm 30 50000 Hz 1 0 1 2 V 0 001 V 5 3 1 kOhm 30 50000 Hz 0 1 15 V 0 01 V 5 3 10 mOhm 30 3000 Hz Presence of the DC offset 5mV will res...

Страница 26: ...0 0 86 0 72 0 57 0 50 0 43 0 36 0 22 470 0 71 0 61 0 50 0 40 0 35 0 30 0 25 0 16 560 0 59 0 51 0 42 0 34 0 30 0 25 0 21 0 13 680 0 49 0 42 0 35 0 28 0 24 0 21 0 17 1000 0 33 0 29 0 24 0 19 0 17 0 14...

Страница 27: ...85 330 0 25 0 17 0 15 0 11 0 10 0 091 0 068 470 0 18 0 14 0 093 0 088 0 084 0 074 1000 0 066 0 063 0 060 0 057 0 054 0 048 2200 0 038 0 036 0 034 0 032 0 031 0 027 3300 0 032 0 030 0 029 0 027 0 026 4...

Отзывы: