Counter-clockwise for 1 beep to reset the counter
•
Counter-clockwise for 1 beep to measure of a positive or negative pulse.
7.8
Signal Generator Mode
Sine, triangle, and rectangle signals are generated. The signal is applied to the device probes. It is
possible to adjust the span (peak-to-peak) of a signal from 0.1 to 2.8 V, a constant bias of ± 0.6 V *, and a
duty cycle of 5 - 95% in 5% increments. Parameters are stored when exiting the mode. On the right of
the screen from top to bottom: sweep, offset, duty cycle and the waveform icon. The frequency in Hz is
displayed in the center of the screen.
The self-shutdown time is 4 times longer than the default in other modes. An external decoupling
capacitor is required to supply a signal to a circuit under voltage. It should be connected to a probe
located closer to the side switch. The other probe is a common one. In this mode, the device does not
turn-off by "flipping" and the function of automatic change of display orientation (left/right hand) is
disabled. It is recommended to use LCR-Reader Kelvin Probe Connector to simplify the device operation
in this mode.
7.8.1
Controls
•
Clockwise for 1 beep to select the item to change.
•
Counter-clockwise for 1 beep to increase the selected parameter.