background image

LCR-Reader-MP Manual

 

Siborg Systems Inc. 

12 

 

7.2.2 

Secondary Parameter 

In Auto mode for inductors quality Q factor is displayed for Q>1 and series Resistance Rs for Q <1, and 
for capacitors, loss tangent for C < 1

μF and ESR 

for C> 1 

μF. Aut

omatic secondary parameter selection 

does not work if the Test Signal is set to Automatic, R is used as the secondary parameter in this case. 

7.2.3 

Test Frequency 

This menu item allows to select the frequency of the test signal. In Auto mode, the device automatically 
selects a frequency of 100Hz, 120 Hz, 1 kHz or 10 kHz, depending on the rating and type of the 
component. Users can also manually select the test frequency. A higher frequency is used for measuring 
small values of capacitances and inductances. Resistor measuring are always made at 1 kHz unless 

RDC+LED

 Mode is selected. If 100 kHz frequency is required, press the jog-wheel and hold for 2 beeps. 

7.2.4 

Test Signal Level 

For more stable readings, select 0.65Vrms or Auto signal level.  Auto is best for measuring loose 
components and in-circuit measurements of R, C, L. When measuring on boards, if at high test signal 
level opens p-n junctions of semiconductors components causing an error in the measurements, then 
the signal level is automatically reduced to 0.1 V peak-to-peak and the value is re-measured. In this case, 
the screen shows the diode symbol and the direction of the p-n junction. 

 

7.2.5 

Period 

Increased measurement time results in more stable readings, reduced time results in faster updates on 
the screen. 

7.2.6 

SER/PAR Mode 

This is 

used to select the equivalent circuit depending on the characteristics of the element. 

7.2.7 

Sound 

The device will beep when the resistance is less than 10 Ohm or a short circuit is found. 

7.2.8 

C-Voltage 

This setting has to be selected in the menu when evaluating electrolytic capacitors for rejection by ESR. 

The voltage is written on the capacitor’s 

housing. 

7.2.9 

Sample Hold 

When selected, the device will hold measurement data on the screen. A beep will sound when the 
measurement is made and the values are held on the screen after the component has been 
disconnected from the device. The device will only reset when a new component is measured or the test 
probes are shorted. 

Содержание LCR-Reader-MP

Страница 1: ......

Страница 2: ...ing off the device 7 6 3 Extended Voltage Range Switch 7 6 4 Controlling device using its orientation 7 6 5 Charging the Battery 7 6 6 Main and System Menu 8 7 Modes of Operation 9 7 1 R L C D Mode 9...

Страница 3: ...mode settings 16 7 6 1 Controls 16 7 7 Frequency Meter mode 16 7 7 1 Changing the mode 16 7 7 2 Frequency Period control 16 7 8 Signal Generator Mode 17 7 8 1 Controls 17 8 Default Settings 18 8 1 De...

Страница 4: ...tion 21 14 4 Details of the Measurement Modes 21 14 4 1 Test Signal Generator 21 14 4 2 Resistance 21 14 4 3 DC Resistance 22 14 4 4 Capacitance 22 14 4 5 DC Capacitance Measurement 23 14 4 6 Inductan...

Страница 5: ...aphics for alternating voltage measurements Jog dial control wheel Multiple Parameter Display primary secondary etc Screen orientation Automatic Manually Sound indication Battery discharge warning ful...

Страница 6: ...onal to the current flowing through the measured component After digitizing the ADC signals the impedance is calculated according to the formula DUT impedance Z Rj Vau Vaj Values of Impedance obtained...

Страница 7: ...the number of pulses from the reference generator and fr is the frequency of the reference generator The principle of measuring the voltage is based on comparing the input signal with the reference vo...

Страница 8: ...the device perpendicular to the floor the screen will flip orientation to the hand being used 2 Turn the probes up to turn on off the indicator light 3 If you place the device on a flat surface with...

Страница 9: ...item by moving the jog wheel in a clockwise or counter clockwise direction and pressing down to select To select the menu item highlight the corresponding menu item and press the jog wheel To set the...

Страница 10: ...the test leads the screen displays information depending on the component and the selected settings in the R L C D mode menu Capacitance Inductance Resistance Diode In automatic mode measurement of c...

Страница 11: ...parasitic capacitance is obtained Please use LCR Reader Calibration Board to set a proper distance between the test leads according to the component size because the parasitic capacitance depends on t...

Страница 12: ...able method of determining the parasitic offset between the tips The dummy PCB uses holes to represent various sizes of components To use the board place the tips into the corresponding holes to the s...

Страница 13: ...components and in circuit measurements of R C L When measuring on boards if at high test signal level opens p n junctions of semiconductors components causing an error in the measurements then the si...

Страница 14: ...electrolytes based on ESR two tables are stored in the memory of the device one for standard aluminum capacitors and the other for Low ESR capacitors Two additional numbers are displayed on the scree...

Страница 15: ...diodes the polarity of the diode the voltage drop in the open state and the reverse current Ir are displayed Diode mode is used for diode parameter extraction using DC bias It is recommended for in ci...

Страница 16: ...to a current and displayed on the screen In the upper line of the screen the resistance is displayed taking into account the input resistance of the device and the lower value of the current To enter...

Страница 17: ...for 2 beeps to automatic selection of T pix V div Press and hold up to 2 beeps to enter the waiting oscilloscope mode In the window set the parameters T pix and V div as indicated above Press and hold...

Страница 18: ...ty cycle and the waveform icon The frequency in Hz is displayed in the center of the screen The self shutdown time is 4 times longer than the default in other modes An external decoupling capacitor is...

Страница 19: ...rms Measurement Period 0 5 s Equivalent Circuit Auto Sound On Data Hold Off 8 2 Default Settings in System Menu The following settings are stored in the device EEPROM when Default in the System Menu i...

Страница 20: ...ht c Use a soft cloth dampened with water to clean the outside surface and cleaning the LCD display device d Do not use liquid solvents or detergents 10 2 Repairs If there is an unexpected measurement...

Страница 21: ...of operation 14 1 Overview R L C Measurement Parameters C R L R R C D C ESR L Q Z Measuring Time 0 25 s 0 5 s 1 0 s 2 0 s Generator Test Signal Frequency 50 Hz 100 Hz 120 Hz 1 kHz 10 kHz 20 kHz 30 kHz...

Страница 22: ...x80x25mm Unit Weight 45 g Weight of the Device and Case 103 g 14 4 Details of the Measurement Modes 14 4 1 Test Signal Generator Test Frequency 50Hz 100Hz 120Hz 1kHz 10kHz 20kHz 30kHz 40kHz 50kHz 60kH...

Страница 23: ...3 5 20M 0 01M 5 5 Calibrate with short and open test leads before measurement Test signal 1 2 V 14 4 4 Capacitance Range Resolution Level Test Frequency 50Hz 100Hz 120Hz 1kHz 10kHz 20kHz 75kHz 100kHz...

Страница 24: ...rement Measurement time 1 sec 14 4 5 DC Capacitance Measurement Range Resolution Error 40 100mF 0 01mF 5 5 680mF 0 1mF 5 5 Discharge capacitors before measurement 14 4 6 Inductance Range Resolution Le...

Страница 25: ...ut Impedance of the Instrument Frequency 75 1000 mV 0 1 mV 5 3 1 kOhm 30 50000 Hz 1 0 1 2 V 0 001 V 5 3 1 kOhm 30 50000 Hz 0 1 15 V 0 01 V 5 3 10 mOhm 30 3000 Hz Presence of the DC offset 5mV will res...

Страница 26: ...0 0 86 0 72 0 57 0 50 0 43 0 36 0 22 470 0 71 0 61 0 50 0 40 0 35 0 30 0 25 0 16 560 0 59 0 51 0 42 0 34 0 30 0 25 0 21 0 13 680 0 49 0 42 0 35 0 28 0 24 0 21 0 17 1000 0 33 0 29 0 24 0 19 0 17 0 14...

Страница 27: ...85 330 0 25 0 17 0 15 0 11 0 10 0 091 0 068 470 0 18 0 14 0 093 0 088 0 084 0 074 1000 0 066 0 063 0 060 0 057 0 054 0 048 2200 0 038 0 036 0 034 0 032 0 031 0 027 3300 0 032 0 030 0 029 0 027 0 026 4...

Отзывы: