Troubleshooting
Inspect the pump again before submitting it for repair. If this does not solve the problem,
contact your Sibata agent.
Symptom Reason
Remedy
No display.
BATTERY LOW is blinking.
Check the DC power supply battery.
The power cord is disconnected
from the main unit.
Connect the power cord to the main
unit, and insert the power plug into
the power outlet.
The pump does not start up.
The settings are wrong.
Make sure that the settings are
correct.
The housing is hot.
Turn the power OFF, and wait for the
housing to cool down.
If this does not remedy the problem,
contact your Sibata agent.
Keys cannot be pressed.
KEY LOCK is lit.
Turn the [POWER] switch to the ON
position.
The flow rate does not decrease. The plug valve has closed.
Open the plug valve.
The inside of the piping between
the sample air inlet port and the
filter holder has become dirty and
clogged.
Check the inside of the piping. If it is
clogged, clean it.
There is a leak between the PM10
impactor and the filter holder.
Connect each part carefully, and
check that there are no leaks.
The backup filter is dirty.
Replace filter element of the backup
filter.
An abnormal ambient
temperature value is indicated.
The connector for the ambient
temperature sensor is not
connected correctly.
Insert the connector for the ambient
temperature sensor all the way in
until you hear a click.
The clock is out of order.
Logging is impossible.
The internal battery is dead.
The battery must be replaced.
Contact your Sibata agent.
"ERROR Set Start Time" is
displayed when the
[START/STOP] key is pressed.
Invalid ON timer setting.
Set the ON timer setting again.
"ERROR Set Stop Time" is
displayed when the
[START/STOP] key is pressed.
Invalid OFF timer setting.
Set the OFF timer setting again.
"ERROR Memory Insufficient" is
displayed when the
[START/STOP] key is pressed.
Insufficient data logger memory.
Clear memory.
"ERROR Memory Insufficient" is
displayed even after executing
Clear Logs.
The total sampling time is set to
130 hours or more.
Set the OFF timers Run Time and
Stop Time so that the total sampling
time is less than 130 hours.
Sampling is canceled before the
OFF timer preset time is
reached and "FAULT" and the
following faulty type are
displayed:
1: Pressure Drop
2: Ambient Temp.
3: Flow Rate
4: Power Low Vol.
1: Filter pressure loss fault
2: Ambient temperature fault
3: Suction flow rate fault
4: Power voltage fault
Remove the cause of the fault stop,
and retry operation.