48
-
F i n i s h e d
P r e s s
[ D O W N ]
t o
d i s p l a y
" T o
t a l
V o l u m e " .
M e m o r y
T i m e
R e m a i n i n g : 1 3 3 h r s
After Sampling is Stopped
Sampling is stopped either by the preset timer
time being reached or by pressing the
[START/STOP] key. When sampling ends
normally, "Finished Memorizing" is displayed,
followed by "Finished...Memory Time
Remaining : ###hrs." ("#" stands for a number.)
"Memory Time Remaining : ###hrs" is the
remaining memory time in which data can be
logged. The maximum data logger time is 133
hours (about 5 days). Also, the maximum
sampling count is 99. When the remaining
memory time has fallen below one hour, or the
sampling count has reached 99, clear memory
(Clear Logs).
(
For details, see P38.)
*
To perform sampling longer than the remaining memory time of the data logger or to
perform sampling with the sampling count exceeding 99, data is not logged in
chronological order at 1-minute intervals.
When "Finished..." is displayed, press the [DOWN] key. The total flow rate during
sampling is displayed.
Sampling Fault
When sampling does not end normally, "FAULT"
is displayed on the 2nd line, and the cause of
the fault stop is displayed on the 3rd line. In the
example on the right, "Pressure Drop" (pressure
loss fault) is displayed.
When a fault stop has occurred, data up to the
point when the fault stop occurred is recorded.
Note, however, that when a power voltage
fault has caused the stop, all data may be
lost depending on the timer setting.
(
For details, see P49.)
List of Fault Stop Items
Displayed
Message
Fault Stop Cause
Pressure Drop
Pressure loss fault:
A pressure loss of 40 kPa or more
continued for 30 seconds.
Ambient Temp.
Ambient temperature fault: An ambient temperature of 45
°
C or
higher continued for 30 seconds.
Flow Rate
Suction flow rate fault:
An error of
±
20 % or more of the set value
continued for one minute.
Power Low Vol.
Power voltage fault:
A power failure occurred or a DC voltage
of 21 V or less continued for 30 seconds.
F i n i s h e d
F A U L T
P r e s s u r e
D
r o p
P r e s s
[ D O W N ]
t o
d i s p l a y
" T o
t a l
V o l u m e " .
M e m o r y
T i m e
R e m a i n i n g : 1 2 9 h r s