25
-
[3] Total Flow Rate Timer Setting (Total Volume)
When "3. Total Time" is selected, the digits
of integer part is displayed highlighted as
shown on the right.
Press the [UP] and [DOWN] keys to
set the value.
When the setting is finished, press the
[ENTER] key.
Next, the digits of fractional part are
displayed highlighted. Set the values in the
same way.
When the setting is finished, press the
[ENTER] key. The newly set value is
entered, and the screen returns to the Setup
sub-menu.
* After setting a new value, if the [MENU/ESCAPE] key is pressed before the
[ENTER] key is pressed, the screen will return to the Setup sub-menu without the
newly entered value being accepted.
Starting Sampling
Perform the following procedure after connecting the filter holder, etc. before the backup
filter.
Press the [MENU/ESCAPE] key several times until the screen returns to the main
screen.
When the screen returns to the main screen, "Setup Set Flow Rate" is initially displayed.
Make sure that the set flow rate is correct.
Pressing the [DOWN] key displays "Setup ON Time" and "Setup OFF Time" in this
order. Make sure that each set value is correct.
If all settings are OK, press the [START/STOP] key. Sampling is started.
When the ON timer is set, the time remaining until sampling is started will be displayed.
When sampling ends successfully, the
screen will change as shown on the right.
Press the [DOWN] key to return to the
main screen.
This completes the series of operations.
For detailed operations and other functions, see the following chapters.
S e t u p
O F F
T i m e
S e
t
T
o
t a
l
V
o
l u m e
m 3
2 0
゜
C
S t a n d a r d
V F
9.
9
9 9
9
9
F i n i s h e d
P r e s s
[ D O W N ]
t o
d i s p l a y
" T o
t a l
V o l u m e " .
M e m o r y
T i m e
R e m a i n i n g : 1 3 3 h r s