Setup – Select Timer (Timer Selection)
When "5 Select Timer" is selected and the
[ENTER] key is pressed, the Select Timer
menu is displayed as shown on the right.
Select the respective menu by the [UP] and
[DOWN] keys, and press the [ENTER] key.
The 3rd and 4th lines show the type of
currently selected ON and OFF timers.
In the example on the right, Delay Time is
selected for the ON timer and Run Time is
selected for the OFF timer.
Select ON Time (ON Timer Selection)
When "1 Select ON Time" is selected and the [ENTER] key is pressed, the Select ON
Time menu is displayed. The default setting is Manual. When an ON timer is selected
from among "Manual," "Delay Time," and "Start Time" by pressing the [UP] and
[DOWN] keys, and the [ENTER] key is pressed, the selection is determined and the
screen returns to the Select Timer menu.
*
After selecting an ON timer in the Select ON Time menu, if the [MENU/ESCAPE] key is
pressed before the [ENTER] key is pressed, the screen will return to the Select Timer
menu without the newly entered value being accepted.
*
When the ON timer is set to "Manual," data is not logged in chronological order at
1-minute intervals. Also, if a power failure occurs during sampling when performing
sampling with the ON timer set to "Manual," all measurement data before the power
failure occurred will be lost. (
For details, see P49.)
Select OFF Time (OFF Timer Selection)
When "2 Select OFF Time" is selected and the [ENTER] key is pressed, the Select
OFF Time menu is displayed. The default setting is Manual. When an OFF timer is
selected from among "Manual," "Run Time," "Stop Time," and "Total Volume" by
pressing the [UP] and [DOWN] keys, and the [ENTER] key is pressed, the selection is
determined and the screen returns to the Select Timer menu.
*
After selecting an OFF timer in the Select OFF Time menu, if the [MENU/ESCAPE] key
is pressed before the [ENTER] key is pressed, the screen will return to the Select Timer
menu without the newly entered value being accepted.
*
When the OFF timer is set to "Manual" or "Total Volume," data is not logged in
chronological order at 1-minute intervals. Also, if a power failure occurs during sampling
when performing sampling with the OFF timer set to "Manual" or "Total Volume," all
measurement data before the power failure occurred will be lost. (
For details, see
P49.)