During Sampling
Processing When a Power Failure Occurs During Pump Operation by 100VAC Power
Power Supply
Fault Stop Setting
OFF Timer Selction
Manual
Run Time
Stop Time
Total Volume
ON
After a
recovery from
a power
failure, the
power voltage
fault is
displayed and
pump
operation is
discontinued.
All
data
before the
power failure
occurred will
be
lost
.
After a
recovery from
a power
failure, the
power voltage
fault is
displayed and
pump
operation is
discontinued.
Data before
the power
failure
occurred will
be
saved
.
After a recovery from a
power failure, the
power voltage fault is
displayed and pump
operation is
discontinued.
Data before the power
failure occurred will be
saved.
After a
recovery from
a power
failure, the
power voltage
fault is
displayed and
pump
operation is
discontinued.
All
data
before the
power failure
occurred will
be
lost
.
* When the ON timer is set to "Manual," after a recovery from a power
failure, the power voltage fault is displayed and pump operation is
discontinued. Also,
all
data before the power failure occurred will be
lost
.
OFF
(Default setting)
After a
recovery from
a power
failure, pump
operation is
continued.
However,
all
data before
the power
failure
occurred will
be
lost
.
After a
recovery from
a power
failure, pump
operation is
continued
until the
remaining
sampling time
reaches 0
(zero). Also,
the data
before the
power failure
occurred will
be
saved
.
When the power failure
is recovered before the
sampling stop time,
pump operation is
continued.
When the power failure
is recovered after the
sampling stop time has
been exceeded, an
OFF timer setting error
is displayed and pump
operation is
discontinued.
In both cases, the data
before the power
failure occurred will be
saved
.
After a
recovery from
a power
failure, pump
operation is
continued
until the total
flow rate
reaches the
set value.
However,
all
data before
the power
failure
occurred will
be
lost
.
* When the ON timer is set to "Manual," after a recovery from a power
failure, the pump returns to the sampling start state and sampling is
repeated from the beginning. Also,
all
data before the power failure
occurred will be
lost
.
*
If a power failure occurs during sampling with a data logger memory error still
displayed, all measurement data before the power failure occurred will be lost after a
recovery from a power failure regardless of the power supply fault stop setting. Also,
the preset flow rate is displayed without the power voltage fault being displayed.