27
-
[2] Main Screen (Standby for Sampling Start)
When the ON timer is set, the sampler stands by for start of sampling (standby for pump
startup). When the [START/STOP] key is pressed in the main screen display, the pump
stands by for startup. Initially, "Waiting Time Remaining" is displayed. The display is
switched to "Waiting Set ON Time" by pressing the [UP] and [DOWN] keys. If the
[START/STOP] key is pressed while the sampler is standing by for start of sampling, the
timer will be canceled. If the timer is canceled before start of sampling, the total flow
rate and other values will remain at their values for the previous measurement.
Main Screen Display (Standby for Sampling Start)
Display Description
Waiting Time Remaining
Displays the time remaining until start of sampling. (
See P43.)
Waiting Set ON Time
Displays the ON timer setting. (
See P43.)
When the [MENU/ESCAPE] key is pressed, the main menu is displayed.
Only "Setup" can be selected. (
See P43.)
Press the [ENTER] key to move to a respective sub-menu. Press the [MENU/ESCAPE]
key to return to the main screen.
[3] Main Screen (During Sampling)
When sampling is started (i.e. pump operation is started), "Sampling Live Flow Rate" is
initially displayed. The display changes in the order shown in the table below by
pressing the [UP] and [DOWN] keys. When the display progresses to "Current Date and
Time," it returns to "Sampling Live Flow Rate. " If the [START/STOP] key is pressed in
the main screen display during sampling, the timer will be canceled and sampling will
be stopped even if timer operation is in progress. Even if sampling is stopped midway,
the measurement values during sampling will be recorded.
Main Screen Display (During Sampling)
Display Description
Sampling Live Flow Rate
Displays the instantaneous flow rate. (
See P44.)
Sampling Total Volume
Displays the total flow rate. (
See P44.)
Sampling Total Samp. Time
Displays the sampling time. (
See P44)
Sampling Time Remaining
Displays the remaining sampling time. (
See P44.)
Current Date and Time
Displays the current date/time. (
See P44.)
When the [MENU/ESCAPE] key is pressed, the main menu is displayed.
To move to a respective sub-menu, press the [UP] and [DOWN] keys to select the item,
and press the [ENTER] key. Press the [MENU/ESCAPE] key to return to the main
screen.
Main Menu Display
Display Description
2 Current Data
Displays the current value. (
See P45.)
3 Setup
Displays the set value. (
See P47.)