During Sampling
Main Screen
[1] Sampling Live Flow Rate (Instantaneous Flow Rate)
The instantaneous flow rate of the pump is displayed. On the last line, for the actual
volume flow rate, "Actual VF" is displayed, and for the volume flow rate after conversion to
the standard state, the standard temperature and "Standard VF" are displayed. (
For
details, see P36.)
[2] Sampling Total Volume (Total Flow Rate)
The total flow rate since sampling was started is displayed. On the last line, for the actual
volume flow rate, "Actual VF" is displayed, and for the volume flow rate after conversion to
the standard state, the standard temperature and "Standard VF" are displayed. (
For
details, see P36.)
[3] Sampling Total Samp. Time (Sampling Time)
The time elapsed since sampling was started is displayed. When "OVER" is displayed on
the 3rd line, this indicates that the time has exceeded 999 hours and 59 minutes.
[4] Sampling Time Remaining (Remaining Sampling Time)
The time remaining until sampling is stopped is displayed. When Total Volume is selected
as the OFF timer, the total flow rate value is displayed. When Manual is selected, "---.---" is
displayed.
[5] Current Date and Time (Current Date/Time)
The current date/time is displayed in the format year/month/day hours:minutes. The year
and time are displayed by the lower two digits of the calendar year and in 24-hour units,