36
-
Setup – Set Flow Unit (Flow Rate Unit Setting)
When "6 Set Flow Unit" is selected and the [ENTER] key is pressed, the Set Flow Unit
menu is displayed. Select the respective menu by the [UP] and [DOWN] keys, and press
the [ENTER] key.
Set Flow Unit Menu Display
Display Description
1 Flow Rate
Sets the suction flow rate unit.
2 Total Volume
Sets the total flow rate unit.
3 Standard Temp.
Sets the standard temperature.
Flow Rate (Suction Flow Rate Setting)
When "1 Flow Rate" is selected and the [ENTER] key is pressed, the Flow Rate menu
is displayed. The default setting is "Actual VF" (actual flow rate unit). When "Standard
VF"(volume flow rate unit after conversion to the standard state) or "Actual VF" is
selected for the suction flow rate unit by pressing the [UP] and [DOWN] keys, and the
[ENTER] key is pressed, the selection is determined and the screen returns to the Set
Flow Unit menu.
*
After selecting the suction flow rate unit in the Flow Rate menu, if the
[MENU/ESCAPE] key is pressed before the [ENTER] key is pressed, the screen will
return to the Set Flow Unit menu without the newly entered value being accepted.
Total Volume (Total Flow Rate Setting)
When "2 Total Volume" is selected and the [ENTER] key is pressed, the Total Volume
menu is displayed. The default setting is "Actual VF" (actual flow rate unit). When
"Standard VF" (volume flow rate unit after conversion to the standard state) or "Actual
VF" is selected for the total flow rate unit by pressing the [UP] and [DOWN] keys, and
the [ENTER] key is pressed, the selection is determined and the screen returns to the
Set Flow Unit menu.
*
After selecting the total flow rate unit in the Total Volume menu, if the
[MENU/ESCAPE] key is pressed before the [ENTER] key is pressed, the screen will
return to the Set Flow Unit menu without the newly entered value being accepted.
Standard Temp. (Standard Temperature Setting)
When "3 Standard Temp." is selected and the [ENTER] key is pressed, the screen
changes to the Standard Temp. menu, and the current standard temperate set value is
displayed. The default setting is 20
°
C. When the standard temperature is selected
from among 0
°
C, 20
°
C and 25
°
C by pressing the [UP] and [DOWN] keys, and the
[ENTER] key is pressed, the selection is determined and the screen returns to the Set
Flow Unit menu.
*
After selecting the standard temperature in the Standard Temp. menu, if the
[MENU/ESCAPE] key is pressed before the [ENTER] key is pressed, the screen will
return to the Set Flow Unit menu without the newly entered value being accepted.