76
Calibrating Moisture Analyzer
s
3
Detailed Measurement
7
When temperature indication
(100 °C) flashes, enter the
calibration temperature.
(Temperature input.)
After 15 minutes has elapsed, the temperature
indication flashes.
Enter the temperature (calibration temperature)
indicated on the thermometer of the temperature
calibration kit.
Press
to start the heating process in high-
temperature range.
*
If it is left while the indicator is
flashing…
Finish entering the figure within
15 minutes after the indicator starts
flashing. If it is not finished, [AboRT]
will be displayed and the calibration
will finish.
8
When temperature indication
(180 °C) flashes, enter the
calibration temperature.
After 15 minutes has elapsed, the temperature
indication flashes.
When [END] is displayed, the temperature
calibration terminates, and the moisture
analyzer returns to the standby state after a
short period of time.
(Temperature input.)
*
If it is left while the indicator is
flashing…
Finish entering the figure within 15
minutes after the indicator starts
flashing. If it is not finished, [AboRT]
will be displayed and the calibration
will finish.
! ! ! !1!0!0
/
/
/
//
//
//
//
/
/
/ /
/ / / / / /
Flashing indicates that the
temperature is stable.
! ! ! ! !!9!5
Enter the temperature indicated on
the thermometer.
! ! ! !1!8!0!
Press
to start the heating
process in high-temperature range.
! ! ! !1!0!0
/
/
/
//
//
//
//
/
/
/ /
/ / / / / /
Flashing indicates that the
temperature is stable.
! ! ! ! !!9!5
Enter the temperature indicated on
the thermometer.
! ! ! !1!8!0!
Press
to start the heating
process in high-temperature range.
! ! ! !1!8!0
/
/
/
//
//
//
//
/
/
/ /
/ / / / / /
Flashing indicates that the
temperature is stable.
! ! ! !1!7!5
Enter the temperature indicated on
the thermometer.
! ! ! !E!N!D
Press
to terminate the heating
process in high-temperature range.
! ! ! !1!8!0
/
/
/
//
//
//
//
/
/
/ /
/ / / / / /
Flashing indicates that the
temperature is stable.
! ! ! !1!7!5
Enter the temperature indicated on
the thermometer.
! ! ! !E!N!D
Press
to terminate the heating
process in high-temperature range.