110
6
tROuBLESHOOtINg guIDE aND OtHER INFORMatION
If You Have a Problem.
If the MOC63u moisture analyzer does not operate normally, check the following points:
Phenomenon
Possible cause
actions taken
See
Nothing appears on the
display.
The power cord is unplugged.
The switchboard is turned off.
The source voltage is incorrect.
The power fuse may be blown
or dismounted.
l
l
l
l
Check the power supply and
source voltage, and connect the
power cord correctly.
Mount the fuse correctly.
l
l
The display does not
change even when
a sample (object) is
placed on the pan.
The sample pan or its supporter
may have become dismounted.
l
Mount the sample pan correctly.
l
Page 35
The display is unstable.
(No stability mark
appears.)
The moisture analyzer is
unstable on the table.
The pan or the sample may
have come in contact with the
windbreak or heater cover.
l
l
Keep the moisture analyzer way
from vibration, wind and other
disturbance.
Place the moisture analyzer on
a stable desk.
l
l
Page 32
The measurement
result is incorrect.
Span calibration has not been
performed.
l
Perform span calibration.
l
Page 40
Zero is not shown before
measurement.
l
Clear the display by pressing
, and start measurement.
l
Page 45
Desired measurement
unit is not shown.
The desired measurement unit
has not yet been set.
l
Set the desired measurement
unit to be displayed.
l
Page 63
The Windows Direct
Function does not start.
Page 84
If you forget your
password
Contact the service representative.
l
The heater does not
light up.
The heater cover may be open.
The heater may be turned off.
The power cable of the heater
may be unplugged.
l
l
l
Close the heater cover.
l
Page 46
n
For further information, see
"If the Windows Direct Function does not work well".
For further information, see
"If the Windows Direct Function does not work well".
troubleshooting
guide