Span calibration can be performed to ensure accurate measurements by the moisture analyzer. Periodical
calibration is recommended to ensure more accurate measurement results.
The following two types of calibrations are provided for the moisture analyzer.
Span calibration
Calibrating the temperature (Option)
Span Calibration
For accurate measurement of the moisture analyzer, span calibration by using a weight.
The calibration result can be recorded if a printer is connected in advance.
^
"Printer Output (Option)", page 91
For further information on the span calibration procedure, in
"The Span Calibration after
Installing the Moisture Analyzer" (
^
page 40)
.
Calibrating the Temperature (Option)
To calibrate the temperature of the MOC63u moisture analyzer, the optional "Temperature
Calibration Kit" is required. For further information on the temperature calibration kit, refer to
the instruction manual attached to the kit.
The calibration record can be output automatically if a printer is connected in advance.
^
"Printer Output (Option)", page 91
1
Insert the thermometer probe
firmly into the temperature
calibration kit.
•
•
n
n
Calibrating
Moisture analyzer