6
Changing Moisture Analyzer Settings
s
3
Detailed Measurement
Either of the following is selectable for the minimum value to be displayed.
For mass: 0.001 grams or 0.01 grams
For others: 0.1 % or 0.01 %
1
Start the menu.
[
]
The menu opens.
2
Select the measurement standard
setting (UNIT) from the menu.
[UNIT]
(The display
currently selected.)
3
Select the type of value to be
referenced.
(Measurement standard selection.)
Display
Description
[
M W
]
Moisture content (Wet Base)
[
D W
]
Dry content (Wet Base)
[
M D
]
Moisture content (Dry Base)
[
W D
]
Dry content (Dry Base)
[
GRAM
]
Mass
u
u