11
Troubleshooting guide
s
6
tROuBLESHOOtINg guIDE aND OtHER INFORMatION
Message code
Possible cause
actions taken
See
ERR.0
The source voltage is abnormal.
l
Unplug the power cord, and
then plug it in again. If the same
message appears again, call a
service representative.
l
Page 38
ERR.C01
ERR.C0
ERR.C0
(Span Calibration error)
A large error in the zero point
or the sensitivity of the pan.
A tare remains on the pan.
The sample pan is not mounted
correctly.
Incorrect weight is placed on
the pan.
l
l
l
l
Press
to return to the
standby state.
Place the correct weight at the
center of the sample pan.
l
Page 40
ERR.oL
ERR.-oL
The oL/-oL status was entered
into during measurement.
l
Check the sample pan.
l
Page 35
oL
-oL
(Overloading)
Too may samples have been
placed.
The sample pan is not mounted
correctly.
l
l
Mount the sample pan correctly.
l
Page 35
TIM.oUT
You started measurement more
than thirty minutes after the zero
point had been adjusted.
l
Stop the measurement by pressing
.
l
Page 45