5
Setting Measuring Conditions
s
3
Detailed Measurement
3
Specify the temperature.
(Temperature input.)
Specify the temperature within the range from
50 °C to 200 °C in 1-degree increments.
4
Specify ∆M.
(∆M input.)
Specify the moisture change rate per 30 seconds
within the range from 0.01 % to 0.1 % in
0.01-percent increments.
Press
to return to step 3.
Press
to save the measurement conditions.
5
Return to the standby state.
Press
to return to the standby state
(measurement display) where measurement can
be performed.
*
When measuring a sample that
has low moisture content...
If you use the standard automatic drying
operation mode to measure a sample
that has a low moisture content (0.1% or
less), it may reach the ending condition
at once and your measurement may
fail. To measure a sample that has a low
moisture content, we recommend that
you use the standard drying time mode
as explained on the next page.