CautION
Instructions
Unplug power cord from receptacle before servicing.
If the power cord is being plugged into receptacle, you may be shocked during servicing.
Prohibitions
Start the service only after the heater and the heater cover have cooled down.
If they are hot, you may burn yourself.
Wipe and clean the main body
with a soft cloth moistened with a
small amount of mild detergent.
Main unit
Do not use organic solvents and
chemicals. Also, do not use a chemical
dust cloth. They may damage the paint
surface and the display panel.
Display panel and keys
You can remove the sample pan
from the moisture analyzer, and
wash it in the water.
Dry the pan well and place it on
the pan supporter.
Sample pan
Carefully remove the glass
case
(
^
page 99)
, and
wipe and clean it with a soft
cloth moistened with a small
amount of mild detergent.
glass case
The windbreak can be detached
from the heater insulate plate
to be washed. Dry it thoroughly
before installing it to the heater
insulate plate.
Windbreak
The heater insulate plate can be
detached from the main unit to be
washed. Dry it thoroughly before
installing it to the main unit.
Heater insulation plate