ZEBV-5084A
7
To avoid electric shock, ensure grounding at power supply ground terminal has a resistance of 100
ohm or less.
To assure performance, keep voltage fluctuation within ±5% and frequency stability within ±0.5 Hz.
In the country where power supply voltage is not AC100V, use a step-down transformer on the
AC100V power-supply line. Refer Fig.5 to install the step-down transformer
<Step-down transformer>
Part number Description
Remarks
225-25315-41
STEP-DOWN TRANSFORMER 1.5KVA
The Step-down transformer and
cables are included.
Please prepare a plug.
*The number
225-25315-41
refers to the Shimadzu Corporation part number.
If power supplies of the above specifications are not available, alterations to room power supply capacity
will be necessary.
Required power supplies for each unit of the iMScope TRIO system are shown in Table 2.
Table2 required power supplies for each unit
Unit
Voltage
Current
Cable length
figure
iMScope TRIO
AC230V 50/60Hz
(single phase)
20 A
10 m
Fig.3
AC100V 50/60Hz
(single phase)
15 A
2.5 m
Fig.4
PC
See the respective instruction manual.
Monitor
Black
・・・・・・
Connects to AC HOT
White
・・・・・・
Connects to AC NEUTRAL
Green / Yellow
・・・
GROUND
Fig.3
Configuration of AC230V power supply cable connection-terminal
(requires connection to circuit breaker or 3-point power plug)
Fig.4 Configuration of AC100V power supply cable terminal
5.3mm dia.
Black
White