2.7 Other items to be prepared
After installation, our field engineers will immediately begin to obtain adjustment acceptance data in
order to confirm normal operation of the unit.
Please prepare the following items in advance as they will be required in the operational check.
(
1
)
Gas
Nitrogen gas N 2
Purity 99% or greater (In case of nitrogen gas generator 97.0 % or greater)
Argon gas Ar
Purity 99.99% or greater
(
2
)
Standard sample for instrument adjustment
Standard sample
225-06613-08 TFANa (sodium trifluroacetate), 200mL
(
3
)
Standard sample for Sensitivity check
Standard sample (will be informed in detail )
To perform data acquisition using a sample prepared by the customer, the following items are required in
addition to the general laboratory apparatuses. Please prepare the following items prior to lectures after
delivery of iMScope TRIO.
If the customer requested to use a sample prepared by the customer during lectures, read “4.2 About
making your samples for iMScope TRIO” before preparing. Preparing multiple glass slides with sections
on them is useful for practicing the spray coating.
(
4
)
Environment for spray coating of the matrix
Airbrush
This is necessary for spray coating of the matrix.
Draft chamber
We recommend you use a draft chamber during the process of
spray coating of the matrix to prevent dispersion of the spray
solution of the matrix.
Dustproof / Gasproof
mask
We recommend you use a Dustproof / Gasproof mask during
the process of spray coating of the matrix to prevent inhaling
the spray solution.
(
5
)
Environment for vapor deposition-coating of the matrix(if matrix vapor deposition equipment is used)
Matrix Vapor Deposition System
This is necessary for vapor depositing of the matrix.
Recommended product : Shimadzu , iMLayer
Table for the vapor deposition
equipment
This needs to be prepared separately.
Draft chamber
If inside of the vapor deposition equipment is contaminated by
the matrix, a large amount of matrix dust may fly up when
the cover is opened. We recommend that the vapor deposition
equipment be installed on the draft chamber to prevent
dispersion of the dust.
Dustproof / Gasproof
mask
If inside of the vapor deposition equipment is contaminated by
the matrix, a large amount of matrix dust may fly up when
the cover is opened. We recommend you use a Dustproof /
Gasproof mask to prevent inhaling the dust.
(
6
)
Consumable parts
Multimeter
This is required to distinguish which is the right side of the
ITO coated glass slide.
Any type of tester will acceptable as
long as it can measure the resistance value.
(Required
measuring range: a few ohms to several megohms)
ITO coated glass slide
Acquisition cannot be properly performed with a normal glass