ZEBV-5084A
5
2250
(
1380
)
Left side view
Open and close view of
the front door during maintenance
Fig.2 Example of installation layout (Left side view)
Table1 Dimensions and Weight
Item
number
Item name
Dimensions
(
mm
)
Weight
W
D
H
(
kg
)
1
iMScope TRIO
1980
1000
1710
900
2
Rotary pump E2M28
597
180
276
40
3
Monitor
See the respective instruction manual.
4
PC
In the example installation given in Fig.1, a separate PC system desk has been prepared, with the
PC system placed on it.
The PC can only be placed on the left side of iMScope TRIO unit.
Install the PC within 0.6 m from the iMScope TRIO.
As the iMScope TRIO unit weights 900 kg, install on a floor that is able to stand load capacities of 900
kg in weight or more.
An additional space of 60cm or more is needed behind and to the right-hand side of the iMScope
TRIO. A further space of 60cm or more is needed to the left of the iMScope TRIO to allow sufficient
space for maintenance. Place equipment that can be readily moved, such as the PC system given in
the example, to the left of the iMScope TRIO unit, and ensure a space of 60cm or more between the
iMScope TRIO unit and walls or other immovable objects.
・
Space behind iMScope TRIO unit
: 60 cm or more
・
Space to right of iMScope TRIO unit
: 60 cm or more
・
Maintenance space to left of iMScope TRIO unit
: 60 cm or more
Front panel may be opened for maintenance. As referring to the figure of the example installation,
place only equipment that can be readily moved on the area where there is a possibility of
interference with the front panel.
Place rotary pump on a firm floor surface at a distance where it can be connected to the iMScope
TRIO unit by attached piping. (Extending piping to a distance greater than this will prevent
assurance of unit performance.) Ensure a distance of 60 cm or more between the back of the iMScope
TRIO and the wall, as the rotary pump will be placed on the floor behind the iMScope TRIO.
It is recommended to place the rotary pump on a stand so that oil change can be easily performed.