Shimadzu iMScope TRIO Скачать руководство пользователя страница 9

 

            ZEBV-5084A

 

OUT

IN

Blue(N)

240V

230V

220V

200V

0

100V

0

a

b

P

S

Brown(L)

Yellow/Green(E)

Blue(N)

Brown(L)

Black

Earth Terminal
(Inside)

To iMScope AC100V in

To power supply

Earth Terminal
(Outside)

225-25316
STEP-DOWN TRNASFORMER 1.5K

225-25317-41
CABLE ASSY, IMSCOPE PS1

225-25319-41
CABLE ASSY, IMSCOPE PS3

225-25318-41
CABLE ASSY, IMSCOPE PS2

225-25320-41
CABLE ASSY, IMSCOPE EARTH

To protective earth

Connects to either 240V, 230,V 220V, or 200V 

terminal according to power supply voltage 

 

Fig.5 How to install the step-down transformer, TBX-1.5KKV 

 

2.5 Gas 

Nitrogen gas (N

2

) and argon gas (Ar) are used for iMScope TRIO. Purities and pressures for each gas are 

as follows: 

Table

3

. Gas purities and pressures 

Gas 

Pressure supplied 

  to iMScope TRIO 

Type 

Purity 

Nitrogen gas

N

2

 

99% or greater 

(In case of nitrogen gas generator 

97.0 % or greater) 

690

800kPa 

Argon gas

A

r)

 

99.99% or greater 

200

500kPa 

 

2.5.1 Nitrogen gas 

 
The method of supplying nitrogen gas, there is a method to use liquid nitrogen, nitrogen gas cylinder, 
nitrogen gas generator. This instrument does not use nitrogen gas other than stopping or adjustment. 
Therefore recommend use the nitrogen gas cylinder or liquid nitrogen. The nitrogen gas generator does 
not absolutely require. Please note that operating hours may be shortened depending on the flow rate of 
the drying gas if using a nitrogen gas cylinder. As an approximate standard, assuming a 47L cylinder 
releasing the gas at a maximum flow rate, this single cylinder would sustain the flow for approximately 5 
hours. (For approximately two times of auto tuning) 

Содержание iMScope TRIO

Страница 1: ...ZEBV 5084A Imaging Mass Microscope iMScope TRIO Pre Installation Requirements Installation Preparations Confirmation...

Страница 2: ...on Copyright 2014 Shimadzu Corporation 2 Information in this manual is subject to change without notice and does not represent a commitment on the part of the vendor 3 Any errors or omissions which ma...

Страница 3: ...t requirements 6 2 4 Required power supply 6 2 5 Gas 7 2 5 1 Nitrogen gas 7 2 5 2 Argon gas 9 2 5 3 General precautions 9 2 5 4 Precautions when using high pressure gas cylinder 10 2 6 Exhaust 10 2 7...

Страница 4: ...reparation required at the customer s site in order to facilitate installation of the iMScope TRIO Please make sure that the content of this document is carefully observed to ensure that the instrumen...

Страница 5: ...crane prepare an approach route and the stopping place of the crane and the carry in entrance to the building Passage and the corners in the building should have the sufficient width and height to mov...

Страница 6: ...space for maintenance Place equipment that can be readily moved such as the PC system given in the example to the left of the iMScope TRIO unit and ensure a space of 60cm or more between the iMScope...

Страница 7: ...on of this instrument must be at least 2 3 m high Including space for working provide a W4 m D3 m H2 3 m site Ventilation This instrument uses flammable solvents for auto tuning and uses large amounts...

Страница 8: ...refers to the Shimadzu Corporation part number If power supplies of the above specifications are not available alterations to room power supply capacity will be necessary Required power supplies for e...

Страница 9: ...es and pressures Gas Pressure supplied to iMScope TRIO Type Purity Nitrogen gas N2 99 or greater In case of nitrogen gas generator 97 0 or greater 690 800kPa Argon gas A 99 99 or greater 200 500kPa 2...

Страница 10: ...Prepare the following pressure regulator and pipes given below or equivalent products if using a nitrogen gas cylinder Pressure regulator Part number Description Requirement Remarks 221 35999 01 Press...

Страница 11: ...en supplying gas by centered gas supply as other instruments being used may be affected by the amount of gas supplied If using a cylinder please use 5m or shorter pipes in general carrier gas piping I...

Страница 12: ...e Draft chamber 20m3 min Be sure to install the oil return kit to suppress the decrease in oil in the rotary pump and clean the exhaust gas from a rotary pump The parts which are require for the exhau...

Страница 13: ...a draft chamber during the process of spray coating of the matrix to prevent dispersion of the spray solution of the matrix Dustproof Gasproof mask We recommend you use a Dustproof Gasproof mask duri...

Страница 14: ...ce For measuring the matrix Pipette and pipette tips 100 1000uL Matrix solvent HPLC grade methanol This is also used for cleaning airbrush Milli Q ultrapure water TFA Vortex mixer For agitating the ma...

Страница 15: ...the following pages are a checklist for each of the installation conditions that have been described in this document Please fill out the checklist yourself and pass it on to the appropriate personnel...

Страница 16: ...Screen to avoid direct contact with airflows from air conditioners Humidity 40 to 70 with no condensation Installation space Space for iMScope TRIO Strength Portable table for PC and monitor Stand fo...

Страница 17: ...e situation 3 point power plug for 230V Power strip for PC Monitor 4 Gas N2 nitrogen gas Purity Supply pressure to iMScope TRIO Distance from gas supply source 99 or greater In case of nitrogen gas ge...

Страница 18: ...us Adjustment acceptance Standard samples for instrument adjustment and sensitivity check are required CD R Blank for installation data record is required TFANa 200mL P N 225 06613 08 Standard sample...

Страница 19: ...cable Strait 2pc Others Consumable parts Multimeter is required to distinguish which is the right side of the ITO coated glass slide ITO coated microscope slides and air duster are consumed regularly...

Страница 20: ...hen affixing the sample Because of the instrument composition we can not analyze that part c Section Thickness To avoid the section damage and the instrument trouble we ask you to slice the tissue wit...

Страница 21: ...al Cutting Temperature compound causes sensitivity decline because the polymer adherents to the sections If you cannot avoid using OCT we ask you to make sections outside the OCT as well If you wish t...

Отзывы: