ZEBV-5084A
11
2.6 Exhaust
The unit emits exhaust gas from the rotary pump and vapor from mobile phase solvents.
Long-term inhalation of these is very harmful to the health.
In order to keep the site clean, it is necessary to direct these gases to a draft chamber or other such
exhaust facility. Five-meter exhaust hoses are supplied with the unit for rotary pump and drain container
discharge. Hoses from the rotary pump and the drain container must be separately connected to exhaust
ducts. A typical draft chamber can be used.
[Example:] Draft chamber: 20m
3
/min.
Be sure to install the oil return kit to suppress the decrease in oil in the rotary pump and clean the
exhaust gas from a rotary pump.
< The parts which are require for the exhaust from a rotary pump>
The following options for the exhaust gas from the rotary pump are available from Shimadzu.
Please specify when ordering.
Part number
Part name
Description
225-05990-92
RP oil return kit
Kit for E2M28
018-31511
SLEEVE,PTFE 15X22 CL
Hoses for E2M28, length 1m
.
Fig6 example of exhaust piping
オイルリターン用
チューブ
オイルミストトラップ(オプション)
廃液キット
ロータリーポンプの排気
ホース(内径φ15X5)
シリコンゴムチューブ(内径φ7X5M)
(廃液キットに付属)
ダクト設備へ
廃液用容器は
お客様準備品
LCMS-IT-TOF
iMScope
Oil mist trap
Exhaust tube from Rotary pump
(ID 15*5)
Tube for oil return
Waste liquid kit
Please prepare
the container
Silicon rubber tube) inner diameter 7 ×5m
(included in Drain kit )
To duct