ZEBV-5084A
17
Item
Condition
Required item
(
* refers to Shimadzu
part number
)
Check
・
Distance from
gas supply sour
・
Must be within 5m from the center
of instrument rear. (Use 5m pipes.)
・
Securing means such
as a gas cylinder holder
(for
two
cylinder
including spare
cylinder)
・
Handle for opening
and closing the gas
cylinder (if the gas
cylinder does not have
a handle)
・
Pressure regulator
:
Same as for N2
nitrogen gas
・
Pipes
:
Same as for N2
nitrogen gas
5.Exhaust
・
Exhaust from
rotary pump
・
Solvent vapor
emission from drain
・
Drying gas
emissions
・
Discharge evacuation gases to a
draft chamber or other such exhaust
facility. (Five-meter exhaust hoses
are supplied with the instrument.
Use of Five-meter hoses is
recommended, however, if it is too
short, prepare longer hoses. )
・
Be sure to install the oil return kit to
rotary pump.
・
Evacuation facilities
such as a draft chamber
・
RP oil return kit
P/N 225-05990-92*
・
Evacuation hose
SLEEVE,PTFE 15X22 CL
P/N 018-31511*
・
Container
□
6.Miscellaneo
us
Adjustment
acceptance
・
Standard samples for instrument
adjustment and sensitivity check are
required.
・
CD-R (Blank) for installation data
record is required.
・
TFANa 200mL
P/N 225-06613-08*
・
Standard sample
(will be informed in
detail )
・
CD-R (Blank)
□
Environment for
spay coating of the
matrix
・
Airbrush is required.
・
Also use of a draft chamber and
Dustproof / Gasproof
mask are
recommended.
・
Airbrush
・
Draft chamber
・
Dustproof / Gasproof
mask
□
Environment for
vapor
deposition-coating
of the matrix
・
Table for the vapor deposition equipment
・
Also use of a draft chamber and
Dustproof / Gasproof
mask are
recommended.
・
Matrix Vapor Deposition
System
(
See the Pre-Installation of
Matrix Vapor Deposition
System
)
・
Draft chamber
・
Dustproof / Gasproof
mask
□
Sample used for
lectures
If the customer requested to use a
sample prepared by the customer
during lectures, read “4.2 About
making your samples for iMScope
TRIO” before preparing.
・
on the ITO-coated
microscope
slide
(Prepared
by
the
customer)
□
Other
required
item
・
Realtime Analysis PC and Postrun
Analysis PC are sold separately.
Please prepare the items we
recommend.
・
One hub and two LAN cables are
・
A set of Realtime
Analysis PC
PC
(
1pc.
)
Display
(
1pc.
)
□