Shimadzu iMScope TRIO Скачать руководство пользователя страница 16

 

            ZEBV-5084A

 

15 

 

Item 

Condition 

Required item 

*  refers  to  Shimadzu 

part number

 

Check 

1.Carry-in 
route 

 

 

The instrument must be movable in 
the building. 

 

(If using a crane) Sufficient space on 
the  carry-in  route,  the  stopping 
place of the crane, and the carry-in 
entrance to the building is available. 

 

□ 

2.Installatin 
room 
 

Temperature 

18 to 28°C 
*at all times, both day and night 
Temperature stability : 
1)

 

+ - 1.5° C    (Term - less than 1° C 
/ 1 Hour) 

2)

 

+ - 0.5° C    (Term - more than 1° 
C / 1 Hour) 

3)

 

+ - 1.5° C    (Term – Rapid change 
within10minutes cycle ) 

 
*Place unit away from direct sunlight 

and in a location where it will not 
come  into  direct  contact  with 
airflows from air conditioners. 

 
*(Reference)  Heat  discharge  of  this 

instrument: around 2,000 kcal/h 

(Depending on the situation) 

Air conditions 

Curtains at the entrance 
of the installation room                       

Screen  to  avoid  direct 

contact  with  airflows 
from air conditioners 

 

□ 

 

Humidity 

40 to 70% (with no condensation) 

 

□ 

 

Installation space 

Space 

for 

iMScope TRIO 

Strength 

 

Portable table for 
PC and monitor 

 
 
 
 

Stand for RP 

 
    W                D              H 
2700mm×1600mm×2300mm 
Ability  to  withstand  weight  loads 
adequately. (900kg) 
Refer to the example installation. The 
table must be wide enough so that 2 
PCs can be placed. 

*Must be moveable 
*The table must be place within 1m 

to the left from the instrument. 

Height can be raised at least 15 cm. 

 
 
 
 
 

Portable  table  for  PC 

and  monitor

Must  be 

movable

 

□ 

 

Miscellaneous 

Ventilation 

 

Interference 

factors

 

 
Must be able to ventilate room well 
 

Dust and fine particles should be minimal. 

Vibration,  magnetic  field  noise,  and 
other 

disturbances 

should 

be 

minimal. 

(Depending on the situation) 

Fan  and/or  ventilation 

opening 

Curtains at the entrance 
of the installation.

 

□ 

3.Power 
supply 

Power supply 
  fluctuation 
(guaranteed 
performance) 

Voltage 

fluctuation

 

Frequency fluctuation

 

 
 
 
 
 

Within ±5% of voltage rating

 

 

50/60Hz, ±0.5Hz

 

 

□ 

Содержание iMScope TRIO

Страница 1: ...ZEBV 5084A Imaging Mass Microscope iMScope TRIO Pre Installation Requirements Installation Preparations Confirmation...

Страница 2: ...on Copyright 2014 Shimadzu Corporation 2 Information in this manual is subject to change without notice and does not represent a commitment on the part of the vendor 3 Any errors or omissions which ma...

Страница 3: ...t requirements 6 2 4 Required power supply 6 2 5 Gas 7 2 5 1 Nitrogen gas 7 2 5 2 Argon gas 9 2 5 3 General precautions 9 2 5 4 Precautions when using high pressure gas cylinder 10 2 6 Exhaust 10 2 7...

Страница 4: ...reparation required at the customer s site in order to facilitate installation of the iMScope TRIO Please make sure that the content of this document is carefully observed to ensure that the instrumen...

Страница 5: ...crane prepare an approach route and the stopping place of the crane and the carry in entrance to the building Passage and the corners in the building should have the sufficient width and height to mov...

Страница 6: ...space for maintenance Place equipment that can be readily moved such as the PC system given in the example to the left of the iMScope TRIO unit and ensure a space of 60cm or more between the iMScope...

Страница 7: ...on of this instrument must be at least 2 3 m high Including space for working provide a W4 m D3 m H2 3 m site Ventilation This instrument uses flammable solvents for auto tuning and uses large amounts...

Страница 8: ...refers to the Shimadzu Corporation part number If power supplies of the above specifications are not available alterations to room power supply capacity will be necessary Required power supplies for e...

Страница 9: ...es and pressures Gas Pressure supplied to iMScope TRIO Type Purity Nitrogen gas N2 99 or greater In case of nitrogen gas generator 97 0 or greater 690 800kPa Argon gas A 99 99 or greater 200 500kPa 2...

Страница 10: ...Prepare the following pressure regulator and pipes given below or equivalent products if using a nitrogen gas cylinder Pressure regulator Part number Description Requirement Remarks 221 35999 01 Press...

Страница 11: ...en supplying gas by centered gas supply as other instruments being used may be affected by the amount of gas supplied If using a cylinder please use 5m or shorter pipes in general carrier gas piping I...

Страница 12: ...e Draft chamber 20m3 min Be sure to install the oil return kit to suppress the decrease in oil in the rotary pump and clean the exhaust gas from a rotary pump The parts which are require for the exhau...

Страница 13: ...a draft chamber during the process of spray coating of the matrix to prevent dispersion of the spray solution of the matrix Dustproof Gasproof mask We recommend you use a Dustproof Gasproof mask duri...

Страница 14: ...ce For measuring the matrix Pipette and pipette tips 100 1000uL Matrix solvent HPLC grade methanol This is also used for cleaning airbrush Milli Q ultrapure water TFA Vortex mixer For agitating the ma...

Страница 15: ...the following pages are a checklist for each of the installation conditions that have been described in this document Please fill out the checklist yourself and pass it on to the appropriate personnel...

Страница 16: ...Screen to avoid direct contact with airflows from air conditioners Humidity 40 to 70 with no condensation Installation space Space for iMScope TRIO Strength Portable table for PC and monitor Stand fo...

Страница 17: ...e situation 3 point power plug for 230V Power strip for PC Monitor 4 Gas N2 nitrogen gas Purity Supply pressure to iMScope TRIO Distance from gas supply source 99 or greater In case of nitrogen gas ge...

Страница 18: ...us Adjustment acceptance Standard samples for instrument adjustment and sensitivity check are required CD R Blank for installation data record is required TFANa 200mL P N 225 06613 08 Standard sample...

Страница 19: ...cable Strait 2pc Others Consumable parts Multimeter is required to distinguish which is the right side of the ITO coated glass slide ITO coated microscope slides and air duster are consumed regularly...

Страница 20: ...hen affixing the sample Because of the instrument composition we can not analyze that part c Section Thickness To avoid the section damage and the instrument trouble we ask you to slice the tissue wit...

Страница 21: ...al Cutting Temperature compound causes sensitivity decline because the polymer adherents to the sections If you cannot avoid using OCT we ask you to make sections outside the OCT as well If you wish t...

Отзывы: