ZEBV-5084A
16
Item
Condition
Required item
(
* refers to Shimadzu
part number
)
Check
Power capacity
・
For iMScope
TRIO
・
For PC, Monitor
・
Ground
(ground resistor)
・
AC230V 20A 50/60Hz
*Avoid sharing power source with
other equipment.
AC100V 15A 50/60Hz
*In the country where power supply
voltage is not AC100V, use a
step-down transformer on the
AC100V power-supply line.
・
Depends on configuration
*Use a power supply equipped with a
ground fault interrupter.
・
Less than 100 ohms
・
Step down transformer :
STEP-DOWN
TRANSFORMER
1.5KVA
P/N 225-25315-41
*
(Depending on the situation)
・
Power plug for Step
Down transformer
・
Extension cable for
100V line
□
Terminals
・
For iMScope
TRIO
・
For PC, Monitor
・
AC230V
:
10m
AC100V
:
2.5m
・
Depends on configuration
(Depending on the situation)
・
3-point power plug for
230V
・
Power strip for PC,
Monitor
□
4.Gas
N
2
(nitrogen gas)
・
Purity
・
Supply pressure
to iMScope TRIO
・
Distance from
gas supply source
・
99% or greater
(In case of nitrogen gas generator
97.0 % or greater)
・
690
~
800kPa
・
Must be within 5m from the center
of instrument rear. (Use 5m pipes.)
(
If using gas cylinders
)
・
Gas cylinder (47L
cylinder
is
recommended )
・
Securing means such
as a gas cylinder holder
(for
two
cylinder
including spare
cylinder)
・
Handle for opening
and closing the gas
cylinder (if the gas
cylinder does not have
a handle)
・
Pressure regulator
:
PPR-N2
P/N 221-35999-01
*
・
Pipes
:
Gas supply
tube, external
P/N 225-09933-91
*
(
If supplying gas
generated from liquid
nitrogen
)
・
Supply through PTFE
pipe with an outer
diameter of 6mm and
an inner diameter of
4mm or larger.
□
Ar (argon gas)
・
Purity
・
Supply pressure
to iMScope TRIO
・
99.99% or greater
・
200
~
500kPa
(
If using gas cylinders
)
・
Gas cylinder ( 3 - 47L
cylinder
is
recommended )
□